Effective radiation dose and eye lens dose in dental cone beam CT: Effect of field of view and angle of rotation

机译:牙锥束CT中的有效辐射剂量和眼镜片剂量:视场和旋转角度的影响

获取原文
获取原文并翻译 | 示例
           

摘要

Objective: To quantify the effect of field of view (FOV) and angle of rotation on radiation dose in dental cone beam CT (CBCT) and to define a preliminary volume-dose model.Methods: Organ and effective doses were estimated using 148 thermoluminescent dosemeters placed in an anthropomorphic phantom. Dose measurements were undertaken on a 3D Accuitomo 170 dental CBCT unit (J. Morita, Kyoto, Japan) using six FOVs as well as fullrotation (360°) and half-rotation (180°) protocols.Results: For the 360° rotation protocols, effective dose ranged between 54 μSv (4×4 cm, upper canine) and 303 μSv (17× 12 cm, maxillofacial). An empirical relationship between FOV dimension and effective dose was derived. The use of a 180° rotation resulted in an average dose reduction of 45% compared with a 360° rotation. Eye lens doses ranged between 95 and 6861 μGy.Conclusion: Significant dose reduction can be achieved by reducing the FOV size, particularly the FOV height, of CBCT examinations to the actual region of interest. Insome cases, a 180° rotation can be preferred, as it has the added value of reducing the scan time. Eye lens doses should be reduced by decreasing the height of the FOV rather than using inferior FOV positioning, as the latter would increase the effective dose considerably.Advances in knowledge: The effect of the FOV and rotation angle on the effective dose in dental CBCT was quantified. The dominant effect of FOV height was demonstrated. A preliminary model has been proposed, which could be used to predict effective dose as a function of FOV size and position.
机译:目的:量化视野(FOV)和旋转角度对牙锥束CT(CBCT)辐射剂量的影响,并定义初步的体积剂量模型。方法:使用148个热致发光剂量计估算器官和有效剂量放在拟人模型中。剂量测量是在3D Accuitomo 170牙科CBCT装置(日本京都J.Morita)上使用六个FOV以及全旋转(360°)和半旋转(180°)方案进行的。结果:对于360°旋转方案有效剂量介于54μSv(4×4 cm,上犬)和303μSv(17×12 cm,颌面部​​)之间。得出了FOV尺寸与有效剂量之间的经验关系。与360°旋转相比,使用180°旋转可使平均剂量减少45%。眼镜片的剂量范围在95到6861μGy之间。结论:通过将CBCT检查的FOV大小(尤其是FOV高度)减小到实际感兴趣的区域,可以显着降低剂量。在某些情况下,最好旋转180°,因为它具有减少扫描时间的附加价值。应通过降低FOV的高度而不是使用劣质FOV定位来降低眼镜片剂量,因为后者会大大增加有效剂量。知识进步:在牙科CBCT中,FOV和旋转角度对有效剂量的影响是量化。证实了FOV高度的主要作用。已经提出了一个初步模型,该模型可以用来预测有效剂量与FOV大小和位置的关系。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号