首页> 外文期刊>British Journal of Radiology >MAPD--an objective way to select mAs for paediatric brain CT.
【24h】

MAPD--an objective way to select mAs for paediatric brain CT.

机译:MAPD-为儿科脑部CT选择mAs的客观方法。

获取原文
获取原文并翻译 | 示例
           

摘要

CT is an advanced imaging modality, but the imaging parameters are normally selected subjectively. For standard head examinations, most of the parameters used are consistent amongst different centres, with the exception of large variations in the selection of the tube current-exposure time product (mAs). As a result, CT images may contain unacceptable levels of noise, or the patient may receive excessive radiation. In this study, the maximum anteroposterior diameter (MAPD) was shown to be a good criterion for mAs selection, and could be measured in a pilot view. 200 paediatric brain CT studies were randomly selected to determine the MAPD at the mid brain level. With knowledge of MAPD distribution, a phantom study was performed to determine the relationship between MAPD and the mAs required for consistent and acceptable image noise. It was found that the required mAs increased linearly with MAPD. Assuming the manufacturer's recommended value is "appropriate" for the average MAPD, the appropriate mAs value could be estimated. Using this method, appropriate mAs values were calculated retrospectively for a group of 240 randomly selected paediatric brain CT studies and compared with the actual mAs subjectively determined by the radiographer. Although their average values were similar, the difference between the calculated and actual values deviated markedly in some cases. When the actual mAs was smaller than the calculated value, higher image noise was observed. However, reduction of image noise was barely observed when the applied mAs was larger than the calculated value. Thus, this method is more objective and appropriate for determination of the mAs value for paediatric brain CT than the traditional subjective method.
机译:CT是一种先进的成像方式,但是成像参数通常是主观选择的。对于标准的头部检查,所使用的大多数参数在不同的中心之间是一致的,除了在管电流暴露时间乘积(mAs)的选择上有很大的差异。结果,CT图像可能包含不可接受的噪声水平,或者患者可能接收到过多的辐射。在这项研究中,最大前后径(MAPD)被证明是选择mAs的良好标准,可以在试验视图中进行测量。随机选择了200例儿科脑CT研究,以确定中脑水平的MAPD。在了解MAPD分布的情况下,进行了幻像研究以确定MAPD和一致且可接受的图像噪声所需的mAs之间的关系。发现所需的mAs随MAPD线性增加。假设制造商对平均MAPD的推荐值“合适”,则可以估算出适当的mAs值。使用这种方法,回顾性地计算了一组240个随机选择的儿科脑CT研究的适当mAs值,并将其与放射线技师主观确定的实际mAs进行了比较。尽管它们的平均值相似,但在某些情况下,计算值与实际值之间的差异明显偏离。当实际mAs小于计算值时,观察到较高的图像噪声。然而,当施加的mAs大于计算值时,几乎观察不到图像噪声的降低。因此,与传统的主观方法相比,该方法更客观,更适合确定小儿脑部CT的mAs值。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号