The aim of this work was to evaluate a new semi-automated intravascular ultrasound (IVUS) border detection method. The method was used to identify the lumen and the external elastic membrane or the borders of stents in 80 IVUS images, randomly selected from 10 consecutive human coronary arteries. These semi-automated results were compared with observations of two experts. Several indices in each case were obtained in order fully to evaluate the method. The time required for identification of the borders was also recorded. The interobserver variability of the method ranged from 1.21% to 5.61%, the correlation coefficient from 0.98 to 0.99, the slope was close to unity (0.94-1.03), the y intercept close to zero and the Williams index value was close to unity (range 0.67-0.91). The time (mean+/-SD) required for the method to identify the borders of the different vessel layers for the whole IVUS sequence was 5.2+/-0.2 min. The results demonstrate that the method is reliable and capable of identifying rapidly and accurately the different vessel layers depicted in IVUS images.
机译:这项工作的目的是评估一种新的半自动血管内超声(IVUS)边界检测方法。该方法用于识别80个IVUS图像中的内腔和外部弹性膜或支架的边界,这些图像是从10个连续的人冠状动脉中随机选择的。将这些半自动化的结果与两位专家的观察结果进行了比较。为了充分评估该方法,分别获得了几种指标。还记录了确定边界所需的时间。该方法的观察者间变异度为1.21%至5.61%,相关系数为0.98至0.99,斜率接近于统一(0.94-1.03),y截距接近于零,Williams指数值接近于统一(范围0.67-0.91)。该方法为整个IVUS序列识别不同血管层边界的方法所需的时间(平均值+/- SD)为5.2 +/- 0.2分钟。结果表明,该方法是可靠的,并且能够快速,准确地识别IVUS图像中描绘的不同血管层。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号