Efficient N-TIMS rhenium isotope measurements on outgassed tantalum filaments: very low filament blanks determined by a 'standard addition' approach

机译:脱气钽丝的高效N-TIMS en同位素测量:通过“标准添加”方法确定极低的丝坯

获取原文
获取原文并翻译 | 示例
           

摘要

We have successfully tested two sets of Ta filaments for their suitability to measure Re isotopes by negative thermal ionization mass spectrometry. No significant Re contribution from the filament material itself could be detected. Therefore Ta filaments are a viable, low cost alternative to high purity Pt filaments which are commonly used for Re-Os isotope analyses. A technique whereby Re is directly loaded with H2SO4 into Ba(NO3)2 and precipitated together with BaSO4 on the filament has resulted in stable and constant ion beams. Furthermore, slight oxidation of the Ta filament surface significantly facilitates point loading in the center of the filament. The BaSO4 loading technique has proved to be less sensitive towards time-consuming pre-analytical heating procedures than Ba(NO3)2 on Pt filaments. Furthermore, Re isotopic ratios remained constant over a wider range of measuring temperatures, in contrast to what we observed during Re analysis from Pt filaments loaded with Ba(NO3)2 or BaSO4.
机译:我们已经成功测试了两组Ta灯丝是否适合通过负热电离质谱法测量Re同位素。丝线材料本身无法检测到明显的Re贡献。因此,Ta灯丝是通常用于Re-Os同位素分析的高纯度Pt灯丝的可行,低成本的替代品。将Re直接与H2SO4一起负载到Ba(NO3)2中并与BaSO4一起沉淀在细丝上的技术产生了稳定且恒定的离子束。此外,Ta细丝表面的轻微氧化显着促进了细丝中心的点负载。事实证明,与在Pt丝上使用Ba(NO3)2相比,BaSO4加载技术对耗时的分析前加热程序不敏感。此外,与我们在对负载Ba(NO3)2或BaSO4的Pt细丝进行Re分析时观察到的相反,Re同位素比率在更宽的测量温度范围内保持恒定。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号