首页> 外文期刊>International journal of mass spectrometry >Comparison of accurate mass LC-MS and MRM LC-MS/MS for the quantification of a therapeutic small interfering RNA
【24h】

Comparison of accurate mass LC-MS and MRM LC-MS/MS for the quantification of a therapeutic small interfering RNA

机译:精确质量LC-MS和MRM LC-MS / MS的比较,用于定量治疗性小分子干扰RNA

获取原文
获取原文并翻译 | 示例
           

摘要

Quantification of oligonucleotide strands from rat plasma has been assessed on two tandem quadrupole systems and a TOF-MS platform following a novel ion-exchange solid-phase extraction and ion-pair chromatographic separation. Limits of quantification are comparable across all three systems with excellent signal-to-noise and accuracy and precision data realised on the TOF-MS to 10 ng/mL. Despite the advances in tandem quadrupole performance in recent years this has not translated into a marked improvement in the detection limits on this instrument geometry. This may prove to be a feature of the tandem MS process for oligonucleotides and could limit the ability to realise lower detection limits on these types of instruments.
机译:在新型离子交换固相萃取和离子对色谱分离之后,已在两个串联四极杆系统和TOF-MS平台上评估了来自大鼠血浆的寡核苷酸链的定量。定量限在所有三个系统上都是可比的,具有出色的信噪比,TOF-MS上达到10 ng / mL的准确度和精密度数据。尽管近年来串联四极杆性能取得了进步,但这并未转化为这种仪器几何形状的检测极限的显着改善。这可能证明是寡核苷酸串联MS过程的一个特征,并且可能限制在这些类型的仪器上实现较低检测限的能力。