Evaluation of a microsphere plate detector for an orthogonal acceleration matrix-assisted laser desorption/ionisation time-of-flight mass spectrometer

机译:正交加速矩阵辅助激光解吸/电离飞行时间质谱仪的微球板检测器评估

获取原文
获取原文并翻译 | 示例
           

摘要

A single 70 mm diameter microsphere plate (MSP) detector was evaluated in an orthogonal acceleration time-of-flight mass spectrometer fitted with a matrix-assisted laser desorption/ionisation source. The ion energy of the spectrometer was approximately 20 kV. The results indicated that the detector performance was excellent, for a single plate device, in terms of background count-rate, temporal characteristics of the pulses and the gains for single ions. A background count rate was observed to be less than 1 count s~(-1) cm~(-2) and the single-ion pulse width was about 800 ps at half height. The plate displayed a gain in the range 1 * 10~6 to 3 * 10~6 for all masses tested. The single ion gain measurements showed no significant decrease with mass while the summed signal intensity (total observed counts) decreased strongly at high mass. The gain distribution data were interpreted in conjunction with this observation to indicate that the detector efficiency was poor for high mass ions.
机译:在配备有基质辅助激光解吸/电离源的正交加速飞行时间质谱仪中评估了单个直径为70毫米的微球板(MSP)检测器。光谱仪的离子能量约为20 kV。结果表明,对于单板设备,在背景计数率,脉冲的时间特性和单离子增益方面,检测器性能非常出色。观察到背景计数率小于1 s s(-1)cm 2(-2),单离子脉冲宽度在半高处约为800 ps。对于所有测试的质量,板显示的增益范围为1 * 10〜6到3 * 10〜6。单个离子增益测量结果显示,质量没有明显降低,而总质量的信号强度(观察到的总计数)在高质量下显着降低。结合这一观察结果解释了增益分布数据,表明对于高质量离子,检测器效率很差。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号