Image quality associated with the use of an MR-compatible incubator in neonatal neuroimaging

机译:与新生儿神经影像学中使用MR兼容培养箱相关的图像质量

获取原文
获取原文并翻译 | 示例
           

摘要

Objectives: MRI in the neonate poses significant challenges associated with patient transport and monitoring, and the potential for diminished image quality owing to patient motion. The objective of this study was to evaluate the usefulness of a dedicated MR-compatible incubator with integrated radiofrequency coils in improving image quality of MRI studies of the brain acquired in term and preterm neonates using standard MRI equipment. Methods: Subjective and objective analyses of image quality of neonatal brain MR examinations were performed before and after the introduction of an MR-compatible incubator. For all studies, the signal-to-noise ratio (SNR) was calculated, image quality was graded (1-3) and each was assessed for image artefact (e.g. motion). Student's t-test and the Mann-Whitney U-test were used to compare mean SNR values. Results: 39 patients were included [mean gestational age 39weeks (range 30-42 weeks); mean postnatal age 13 days (range 1-56 days); mean weight 3.5 kg (range 1.4-4.5 kg)]. Following the introduction of the MR-compatible incubator, diagnostic quality scans increased from50 to 89%and motion artefact decreased from73 to 44%of studies. SNR did not increase initially, but, when using MR sequences and parameters specifically tailored for neonatal brain imaging, SNR increased from 70 to 213 (p=0.001). Conclusion: Use of an MR-compatible incubator in neonatal neuroimaging provides a safe environment for MRI of the neonate and also facilitates patient monitoring and transport. When specifically tailored MR protocols are used, this results in improved image quality.
机译:目标:新生儿的MRI带来了与患者运输和监测相关的重大挑战,以及由于患者运动而导致图像质量下降的可能性。这项研究的目的是评估带有集成射频线圈的专用MR兼容培养箱在提高使用标准MRI设备采集的足月和早产儿的MRI图像质量方面的有用性。方法:在引入MR兼容培养箱之前和之后,对新生儿脑部MR检查的图像质量进行了主观和客观分析。对于所有研究,均计算信噪比(SNR),将图像质量分级(1-3),并对每个图像的伪像(例如运动)进行评估。使用学生t检验和Mann-Whitney U检验比较平均SNR值。结果:纳入39例患者[平均胎龄39周(范围30-42周);平均出生后13天(1-56天);平均重量3.5公斤(范围为1.4-4.5公斤)]。引入MR兼容培养箱后,研究的诊断质量扫描从50%提高到89%,运动伪影从73%降低到44%。 SNR最初并没有增加,但是当使用专门为新生儿脑成像量身定制的MR序列和参数时,SNR从70增加到213(p = 0.001)。结论:在新生儿神经影像学中使用兼容MR的培养箱可为新生儿MRI提供安全的环境,并有助于患者监测和运输。当使用专门定制的MR协议时,这会提高图像质量。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号