Analysis of metastable decay by time-of-flight coincidence and kinetics energy measurements

机译:通过飞行时间重合和动能测量来分析亚稳态衰变

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

A procedure for fragmentation pattern analysis of metastable ions, emitted in photon or ion induced desorption processes, is described. The method is based on a combination of three well-established techniques, namely: time-of-flight (TOF), coincidence counting, and electrostatic energy analysis in reflectron mass spectrometers. The key improvement comes from a judicious selection of the electrostatic mirror voltages for running the coincidence measurements in the best signaloise conditions. The TOFs of stable species, as well as those of the neutral and ion fragments produced by all expected unimolecular-binary decays occurring in the field-free region before the mirror, are calculated and plotted as a function of the mirror voltage. The mirror voltage for coincidence measurements are selected after comparison of these dynamics plots of reflected and nonreflected species with experimental noncoincidence TOF spectra. To illustrate this procedure, the (LiF)_nLi~(+*) → (LiF)_(n-m)Li~+ + (LiF)_m~0 fragmentation pattern is analyzed.
机译:描述了一种在光子或离子诱导的解吸过程中发射的亚稳态离子的碎片图谱分析程序。该方法基于三种公认的技术的结合,即:飞行时间(TOF),同时计数和反射电子质谱仪中的静电能分析。关键的改进来自明智选择的静电镜电压,以便在最佳信号/噪声条件下运行同时测量。计算稳定物种的TOF,以及由镜前无场区域中发生的所有预期的单分子二元衰变产生的中性和离子碎片的TOF,并将其作为镜电压的函数进行绘制。在将反射和未反射物种的这些动态图与实验非符合TOF光谱进行比较之后,选择用于符合性测量的镜像电压。为了说明此过程,分析了(LiF)_nLi〜(+ *)→(LiF)_(n-m)Li〜+ +(LiF)_m〜0碎片模式。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号