Estimating the effective dose to children undergoing heart investigations--a phantom study.

机译:估算进行心脏检查的儿童的有效剂量-一项幻像研究。

获取原文
获取原文并翻译 | 示例
           

摘要

The aim of the investigation was to assess the conversion factor (F) for derivation of effective dose from measured dose-area product (DAP) during radiological examination of congenital heart diseases. Two anthropomorphic phantoms corresponding to a 1-year-old and a 5-year-old child were irradiated at several projections to imitate irradiation conditions at heart examinations. Organ doses were measured using thermoluminescent dosimeters for calculation of mean organ doses and effective dose according to ICRP. DAP values were measured simultaneously. The conversion factor (F) was calculated from the ratio of effective dose (mSv) to DAP value (Gy cm2). The conversion factor (F) correlated strongly to the size of the phantom but less to the irradiation projection. However, at major beam angulation and at lateral projection F deviated note-worthily from that obtained at true or slightly angulated frontal views. Effective dose can therefore be estimated from the DAP values at heart investigation using two different F values. The following values are recommended for PA and lateral view respectively: for children weighing 7-11 kg, 1.8 and 1.4; for children weighing 15-26 kg 0.9 and 0.7.
机译:研究的目的是评估先天性心脏病放射检查期间从测量的剂量面积积(DAP)得出有效剂量的转换因子(F)。在几个投影上照射了分别对应1岁和5岁孩子的两个拟人化体模,以模仿心脏检查时的照射条件。使用热辐射剂量计测量器官剂量,以根据ICRP计算平均器官剂量和有效剂量。同时测量DAP值。由有效剂量(mSv)与DAP值(Gy cm2)之比计算转化因子(F)。转换因子(F)与体模的大小密切相关,而与辐射投影的相关性较小。但是,在主光束角度和侧向投影F时,与在真实或略有角度的正视图中获得的值明显不同。因此,可以在心脏检查时使用两个不同的F值从DAP值估算有效剂量。对于PA和侧视图,建议分别使用以下值:体重7-11公斤,1.8和1.4的儿童;适用于体重15-26公斤的儿童0.9和0.7。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号