首页> 外文期刊>International journal of laboratory hematology >A study of atypical APTT derivative curves on the ACL TOP coagulation analyser.
【24h】

A study of atypical APTT derivative curves on the ACL TOP coagulation analyser.

机译:在ACL TOP凝血分析仪上研究非典型APTT导数曲线。

获取原文
获取原文并翻译 | 示例
       

摘要

INTRODUCTION: The graphical representation of clotting data can provide useful information. A novel feature of the ACL TOP software allows display of clot reaction curves with superimposed first and second derivative curves. We noted atypical derivative plots associated with normal 'S'-shaped clot reaction curves in some abnormal activated partial thromboplastin time (APTT). METHODS: We tested 179 citrate plasmas with four different APTT reagents on the ACL TOP. We documented the prevalence and utility of these APTT atypical derivative curves in these patients. RESULTS: Ellagic acid APTT reagents commonly produce atypical derivative curves despite normal 'S'-shaped clot reaction curves. Occasionally, these atypical second derivative curves may cause incorrect APTT results. With silica activator APTT reagents, atypical derivative curves were associated with genuine coagulation abnormalities such as single factor deficiencies and lupus anticoagulants (LAs). To our knowledge, this novel finding has not been reported. CONCLUSION: Atypical derivative curves seen in ellagic acid APTTs are of limited diagnostic use because of the frequency with which they occur. This may be related to the need to optimize the data reduction utilized on the ACL TOP for these reagents. With silica activator APTTs, the presence of atypical derivative curves proved to be a very simple tool when troubleshooting unexpected abnormal APTT results, commonly predicting a factor deficiency or LA that would warrant further investigation. The cause of these aberrant derivative curves is probably related to abnormal thrombin generation in the APTT test and warrants further study.
机译:简介:凝血数据的图形表示可以提供有用的信息。 ACL TOP软件的新颖功能允许显示凝块反应曲线以及叠加的一阶和二阶导数曲线。我们注意到在某些异常激活的部分凝血活酶时间(APTT)中与正常的“ S”形血块反应曲线相关的非典型导数图。方法:我们在ACL TOP上用四种不同的APTT试剂测试了179种柠檬酸盐血浆。我们记录了这些患者中这些APTT非典型导数曲线的普遍性和实用性。结果:尽管正常的“ S”形凝块反应曲线正常,但鞣花酸APTT试剂通常会产生非典型的导数曲线。有时,这些非典型的二阶导数曲线可能会导致错误的APTT结果。使用二氧化硅活化剂APTT试剂,非典型导数曲线与真正的凝血异常相关,例如单因素缺陷和狼疮抗凝剂(LAs)。据我们所知,这一新颖发现尚未得到报道。结论:鞣花酸APTTs中发现的非典型导数曲线由于其出现的频率而在诊断上用途有限。这可能与优化这些试剂在ACL TOP上利用的数据缩减的需求有关。对于二氧化硅活化剂APTT,当对意外的异常APTT结果进行故障诊断时,非典型导数曲线的存在被证明是一种非常简单的工具,通常预测因子缺乏症或LA值得进一步研究。这些异常导数曲线的原因可能与APTT测试中凝血酶异常产生有关,值得进一步研究。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号