Synthetic isotope mixtures for the calibration of isotope amount ratio measurements of carbon

机译:合成同位素混合物,用于校准碳的同位素量比

获取原文
获取原文并翻译 | 示例
           

摘要

Synthetic isotope mixtures for the calibration of carbon isotope amount ratio measurements have been prepared by mixing carbon tetrafluoride highly enriched in ~(13)C with carbon tetrafluoride depleted in ~(13)C. Mixing procedures based on volumetry and gravimetry are described. The mixtures served as primary measurement standards for the calibration of isotope amount ratio measurements of the Isotopic Reference Materials PEF1, NBS22 and USGS24. Thus SI-traceable measurements of absolute carbon isotope amount ratios have been performed for the first time without any hypothesis needed for a correction of oxygen isotope abundances, such as is the case for measurements on carbon dioxide. As a result, "absolute" carbon isotope amount ratios determined via carbon tetrafluoride have smaller uncertainties than those published for carbon dioxide. From the measurements of the Reference Materials concerned, the absolute carbon isotope amount ratio of Vienna Pee Dee Belemnite (VPDB)-the hypothetical material upon which the scale for relative carbon isotope ratio measurements is based-was calculated to be R_(13)(VPDB) = (11 101 ± 16) * 10~(-6).
机译:通过将高度富含〜(13)C的四氟化碳与耗尽~~(13)C的四氟化碳混合,可以制备用于校准碳同位素量比测量的合成同位素混合物。描述了基于体积和重量分析的混合程序。混合物用作同位素参考物质PEF1,NBS22和USGS24同位素量比测量值校准的主要测量标准。因此,首次进行了可溯源到SI的绝对碳同位素含量比的测量,而无需进行任何假设来校正氧同位素丰度,例如对二氧化碳的测量。结果,通过四氟化碳测定的“绝对”碳同位素量比的不确定度比对二氧化碳公布的不确定度小。从相关参考材料的测量中,维也纳皮伊迪贝拉米特(VPDB)的绝对碳同位素含量比(假设用于相对碳同位素比测量的标度的材料)计算为R_(13)(VPDB )=(11101±16)* 10〜(-6)。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号