Variability of dental cone beam CT grey values for density estimations

机译:牙锥束CT灰度值的可变性以进行密度估算

获取原文
获取原文并翻译 | 示例
           

摘要

Objective: The aim of this study was to investigate the use of dental cone beam CT (CBCT) grey values for density estimations by calculating the correlation with multislice CT (MSCT) values and the grey value error after recalibration. Methods: A polymethyl methacrylate (PMMA) phantom was developed containing inserts of different density: air, PMMA, hydroxyapatite (HA) 50mgcm~(-3), HA 100, HA 200 and aluminium. The phantom was scanned on 13 CBCT devices and 1 MSCT device. Correlation between CBCT grey values and CT numbers was calculated, and the average error of the CBCT values was estimated in the medium-density range after recalibration.Results: Pearson correlation coefficients ranged between 0.7014 and 0.9996 in the full-density range and between 0.5620 and 0.9991 in the medium-density range. The average error of CBCT voxel values in the medium-density range was between 35 and 1562.Conclusion: Even though most CBCT devices showed a good overall correlation with CT numbers, large errors can be seen when using the grey values in a quantitative way. Although it could be possible to obtain pseudo-Hounsfield units from certain CBCTs, alternative methods of assessing bone tissue should be further investigated. Advances in knowledge: The suitability of dental CBCT for density estimations was assessed, involving a large number of devices and protocols. The possibility for grey value calibration was thoroughly investigated.
机译:目的:本研究的目的是通过计算与多层CT(MSCT)值的相关性以及重新校准后的灰度值误差,来研究使用牙锥束CT(CBCT)灰度值进行密度估计。方法:开发了一种聚甲基丙烯酸甲酯(PMMA)体模,其中包含不同密度的插入物:空气,PMMA,羟基磷灰石(HA)50mgcm〜(-3),HA 100,HA 200和铝。幻像在13个CBCT设备和1个MSCT设备上进行了扫描。重新校准后,计算了CBCT灰度值与CT值之间的相关性,并估计了中密度范围内CBCT值的平均误差。结果:皮尔逊相关系数在全密度范围内介于0.7014至0.9996之间,而在0.5620至0.9620之间。在中密度范围内为0.9991。在中等密度范围内,CBCT体素值的平均误差在35到1562之间。结论:尽管大多数CBCT器件与CT值显示出良好的整体相关性,但在定量使用灰度值时仍可看到较大的误差。尽管可以从某些CBCT中获得伪Hounsfield单位,但应进一步研究评估骨组织的替代方法。知识进步:评估了牙科CBCT对密度估计的适用性,涉及大量设备和协议。灰度值校准的可能性已被彻底研究。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号