首页> 外文期刊>British Journal of Radiology >Radiation dose evaluation in 64-slice CT examinations with adult and paediatric anthropomorphic phantoms.
【24h】

Radiation dose evaluation in 64-slice CT examinations with adult and paediatric anthropomorphic phantoms.

机译:成人和儿童拟人化体模的64层CT检查中的辐射剂量评估。

获取原文
获取原文并翻译 | 示例
           

摘要

The objective of this study was to evaluate the organ dose and effective dose to patients undergoing routine adult and paediatric CT examinations with 64-slice CT scanners and to compare the doses with those from 4-, 8- and 16-multislice CT scanners. Patient doses were measured with small (<7 mm wide) silicon photodiode dosemeters (34 in total), which were implanted at various tissue and organ positions within adult and 6-year-old child anthropomorphic phantoms. Output signals from photodiode dosemeters were read on a personal computer, from which organ and effective doses were computed. For the adult phantom, organ doses (for organs within the scan range) and effective doses were 8-35 mGy and 7-18 mSv, respectively, for chest CT, and 12-33 mGy and 10-21 mSv, respectively, for abdominopelvic CT. For the paediatric phantom, organ and effective doses were 4-17 mGy and 3-7 mSv, respectively, for chest CT, and 5-14 mGy and 3-9 mSv, respectively, for abdominopelvic CT. Doses to organs at the boundaries of the scan length were higher for 64-slice CT scanners using large beam widths and/or a large pitch because of the larger extent of over-ranging. The CT dose index (CTDI(vol)), dose-length product (DLP) and the effective dose values using 64-slice CT for the adult and paediatric phantoms were the same as those obtained using 4-, 8- and 16-slice CT. Conversion factors of DLP to the effective dose by International Commission on Radiological Protection 103 were 0.024 mSvmGy(-1)cm(-1) and 0.019 mSvmGy(-1)cm(-1) for adult chest and abdominopelvic CT scans, respectively.
机译:这项研究的目的是评估接受64层CT扫描仪进行常规成人和儿科CT检查的患者的器官剂量和有效剂量,并将其与4层,8层和16层CT扫描仪的剂量进行比较。用小的(小于7毫米宽)硅光电二极管剂量计(总共34个)测量患者的剂量,将其植入成人和6岁儿童拟人体模中的各种组织和器官位置。在个人计算机上读取光电二极管剂量计的输出信号,从中计算出器官和有效剂量。对于成人体模,胸部CT的器官剂量(对于扫描范围内的器官)和有效剂量分别为8-35 mGy和7-18 mSv,腹部腹腔的器官剂量分别为12-33 mGy和10-21 mSv CT。对于小儿体模,胸部CT的器官和有效剂量分别为4-17 mGy和3-7 mSv,而腹部盆腔CT的器官和有效剂量分别为5-14 mGy和3-9 mSv。对于64排CT扫描仪,使用较大的光束宽度和/或较大的间距时,在扫描长度边界处对器官的剂量较高,这是因为超量程的程度较大。成人和儿童体模的CT剂量指数(CTDI(vol)),剂量长度乘积(DLP)和有效剂量值与使用4层,8层和16层体模获得的CT剂量指数相同CT。对于成人胸部和腹部盆腔CT扫描,DLP转换为国际放射防护委员会103的有效剂量的转换因子分别为0.024 mSvmGy(-1)cm(-1)和0.019 mSvmGy(-1)cm(-1)。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号