Dose verification of intensity-modulated arc therapy using an ERGO++ treatment planning system and Elekta internal multileaf collimators for prostate cancer treatment.

机译:使用ERGO ++治疗计划系统和Elekta内部多叶准直器对前列腺癌进行强度调制电弧治疗的剂量验证。

获取原文
获取原文并翻译 | 示例

摘要

Dose verification of intensity-modulated arc therapy using an ERGO++ treatment planning system and Elekta internal multileaf collimators is described. Prostate intensity-modulated arc therapy was planned using the arc modulation optimization algorithm inverse planning module of ERGO++. After transferring the plan to Elekta Synergy's controller (Elekta Ltd, Crawley, UK), the isocentre dose was measured and compared with a calculated dose using a pinpoint chamber and a water phantom in a cylindrical acrylic enclosure. Subsequently, an EDR2 film was placed inside a multilayer plastic phantom, and total dose distributions were measured in three axial planes as well as in the coronal and sagittal planes to compare the actual dose with the calculated dose. The dose discrepancy at the isocentre was 1.7%. The calculated gamma indices were less than 1 over 90% of the three axial planes, as well as in the coronal and sagittal planes, having a dose greater than 50% of the maximum target dose.
机译:描述了使用ERGO ++治疗计划系统和Elekta内部多叶准直仪进行强度调制电弧治疗的剂量验证。使用ERGO ++的电弧调制优化算法逆向规划模块来规划前列腺强度调制的电弧疗法。将计划转移到Elekta Synergy的控制器(Elekta Ltd,Crawley,英国)之后,测量了等中心剂量,并使用精确定位室和水模体在圆柱形丙烯酸外壳中与计算出的剂量进行了比较。随后,将EDR2膜放置在多层塑料体模内,并在三个轴向平面以及冠状面和矢状面中测量总剂量分布,以将实际剂量与计算出的剂量进行比较。等中心点的剂量差异为1.7%。在三个轴向平面的90%以上以及在冠状平面和矢状平面中,所计算的伽玛指数小于1,且其剂量大于最大目标剂量的50%。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号