首页> 外文期刊>International journal of laboratory hematology >An evaluation of the SPIFE 3000 semi-automated gel electrophoresis system for the identification of hemoglobin variants and comparison of relative electrophoretic mobilities with manual gel electrophoresis methods.
【24h】

An evaluation of the SPIFE 3000 semi-automated gel electrophoresis system for the identification of hemoglobin variants and comparison of relative electrophoretic mobilities with manual gel electrophoresis methods.

机译:评估SPIFE 3000半自动凝胶电泳系统以鉴定血红蛋白变体并通过手动凝胶电泳方法比较相对电泳迁移率。

获取原文
获取原文并翻译 | 示例
           

摘要

Laboratory identification of hemoglobin (Hb) variants can involve multiple techniques. The use of semi-automated instruments that perform gel electrophoresis and staining, such as the SPIFE 3000 electrophoresis system, can greatly reduce the labor required for these commonly used techniques. We performed a comparison of the method involved in SPIFE 3000 system with those of manual gel electrophoresis. A total of 22 540 samples were analyzed using the SPIFE 3000, and compared with mobilities on cellulose acetate and citrate agar gels using standard manual methods. The results were compared using relative electrophoretic mobilities (REM). Of the 191 Hb variants identified, only 13 had REM that differed from manual electrophoresis when analyzed using the SPIFE 3000 system. One variant (Hb O-Indonesia) showed different mobility on both acid and alkaline gels, two (Hb E, Hb Sunshine Seth) on alkaline gel only, and 10 (Hbs N-Baltimore, N-Seattle, O-Arab, Shelby, Summer Hill, Tak, Hasharon, M-Iwate, Q-Iran, and Setif) on acid gels only. The SPIFE 3000 semi-automated electrophoresis system produces similar results when compared with those of standard manual electrophoresis methods.
机译:实验室鉴定血红蛋白(Hb)变异可能涉及多种技术。使用执行凝胶电泳和染色的半自动化仪器,例如SPIFE 3000电泳系统,可以大大减少这些常用技术所需的劳动。我们将SPIFE 3000系统中涉及的方法与手动凝胶电泳中的方法进行了比较。使用SPIFE 3000对总共22 540个样品进行了分析,并使用标准手动方法与在醋酸纤维素和柠檬酸琼脂凝胶上的迁移率进行了比较。使用相对电泳迁移率(REM)比较结果。在使用SPIFE 3000系统进行分析时,在鉴定出的191种Hb变体中,只有13种具有与手动电泳不同的REM。一种变体(Hb O-印度尼西亚)在酸性和碱性凝胶上均显示出不同的迁移率,两种(Hb E,Hb阳光Seth)仅在碱性凝胶上显示,而10种(Hbs N-巴尔的摩,N-Seattle,O-Arab,Shelby, Summer Hill,Tak,Hasharon,M-Iwate,Q-Iran和Setif)只能在酸性凝胶上使用。与标准手动电泳方法相比,SPIFE 3000半自动电泳系统产生相似的结果。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号