Considerations on the measurement of practical peak voltage in diagnostic radiology.

机译:诊断放射学中实际峰值电压的测量注意事项。

获取原文
获取原文并翻译 | 示例
           

摘要

An evaluation of the non-invasive measurement of the practical peak voltage (PPV) in the quality control of X-ray units used in diagnostic radiology was carried out. Two instruments were employed: the PTW Diavolt Universal Tester with readings in PPV and the Waveform Tester for X-rays (WATEX) prototype proposed here, which uses a PIN structure (P-type diffusion, Intrinsic region, N-type diffusion) photodiode as a sensor. The reference for the measurements was the voltage signal obtained in an oscilloscope from an invasive high voltage divider in order to verify the accuracy and precision of the measurements. The readings of the PPV in the Diavolt show a systematic error between 1% and 8%, always being less than the real value. An explanation for this difference is proposed, based on the relation between the effect of the X-rays in the film and the response of the sensor to the product of the applied voltage to the X-ray tube (peak voltage kVp), and the anode current. This explanation was confirmed using the WATEX waveform tester.
机译:对用于诊断放射学的X射线装置的质量控制中的实际峰值电压(PPV)的非侵入性测量进行了评估。使用了两种仪器:在PPV中具有PPV读数的PTW Diavolt通用测试仪和此处提出的X射线波形测试仪(WATEX)原型,其使用PIN结构(P型扩散,本征区域,N型扩散)光电二极管作为一个传感器。测量的参考是在示波器中从有创高压分压器获得的电压信号,以验证测量的准确性和精度。 Diavolt中PPV的读数显示系统误差在1%到8%之间,始终小于实际值。根据胶片中X射线的效果与传感器对X射线管施加的电压乘积(峰值电压kVp)的响应之间的关系,提出对此差异的解释。阳极电流。使用WATEX波形测试仪证实了这一解释。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号