Patient dose optimization in plain radiography based on standard exposure factors.

机译:基于标准暴露因素的普通X射线摄影中的患者剂量优化。

获取原文
获取原文并翻译 | 示例
           

摘要

A computational technique for assessing patient dose in plain radiography is described allowing a large number of examinations to be assessed and enabling dose optimization to be promoted. Entrance surface dose (ESD) was calculated for more than 1500 standard exposure settings in an initial dose assessment. Validation of the technique showed good agreement with thermoluminescent dosimetry and showed broad agreement between the standard exposures and the exposure settings used in practice. The dose assessment was repeated 18 months later using the same techniques for almost 2000 standard exposure settings. In both cases, calculated doses showed good compliance with national diagnostic reference levels where available. Suggested investigation levels were established and set at twice the mean dose for each of 47 examinations. Radiology departments were encouraged to review and optimize doses exceeding these levels. The computed mean ESD in the review study was less than the corresponding value in the initial study in 37 of the 47 examinations. The dose reduction was attributable partly to equipment replacement, but primarily to optimization of exposure settings. The technique employed here provides a valid and cost effective method of complying with statutory requirements for the assessment of representative patient dose and is useful in assisting the ongoing process of dose optimization.
机译:描述了一种用于在普通放射线照相术中评估患者剂量的计算技术,该技术允许评估大量检查并促进剂量优化。在初始剂量评估中,针对超过1500种标准暴露设置计算了入口表面剂量(ESD)。该技术的验证显示了与热发光剂量法的良好一致性,并显示了标准曝光量和实际使用的曝光量设置之间的广泛一致性。 18个月后,使用相同的技术针对近2000个标准暴露设置重复进行剂量评估。在这两种情况下,计算的剂量均符合国家诊断参考水平(如可用)。建立了建议的调查水平,并将其设置为47项检查的平均剂量的两倍。鼓励放射科检查并优化超过这些水平的剂量。在47项检查中的37项检查中,审查研究中计算出的平均ESD小于初始研究中的相应值。剂量减少部分归因于设备更换,但主要归因于暴露设置的优化。此处采用的技术提供了一种符合法定要求以评估代表性患者剂量的有效且具有成本效益的方法,可用于协助进行中的剂量优化过程。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号