Validation of the international normalized ratio (INR) in a new point-of-care system designed for home monitoring of oral anticoagulation therapy.

机译:在设计用于家庭口服抗凝治疗监测的新型即时护理系统中,对国际标准化比率(INR)的验证。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

The new CoaguChek XS system is designed for use in patient self testing with a measuring range from 0.8 INR up to 8.0 INR, which has been calibrated against the mean INR of rTF/95 and ERM-AD149. This study was performed to confirm the correct INR results received from two routinely manufactured lots of test strips when compared with the international reference preparations (IRP) rTF/95 and ERM-AD149. At one study site capillary and noncitrated venous whole blood samples from 20 normal donors and 62 anticoagulated patients were applied to two test strip lots of the new system in duplicate. Additionally blood was collected in citrate tubes, processed to plasma, and PT results were obtained using rTF/95 and ERM-AD149 by the manual tilt tube method. Method comparisons of the INR results of the CoaguChek XS system vs. the mean INR of the IRP demonstrated a mean relative bias of -0.02% to -0.4%, mean absolute relative deviations of 6.4-9.6%, and accuracy observing >95% of CoaguChek XS INR within limits of +/-14% to +/-21.5% to the mean INR of the IRP. The results of the study confirm the successful calibration of two lots of the new CoaguChek XS system, demonstrate the validity of the calibration concept and prove the accuracy of the new system in comparison with the IRP. Clinical decisions in oral anticoagulation therapy may be reliably made upon the INR results of the new system.
机译:新的CoaguChek XS系统设计用于患者自检,测量范围从0.8 INR到8.0 INR,已针对rTF / 95和ERM-AD149的平均INR进行了校准。与国际参考制剂(IRP)rTF / 95和ERM-AD149相比,进行这项研究是为了确认从两个常规制造的试纸中获得正确的INR结果。在一个研究地点,将来自20位正常供体和62位抗凝患者的毛细管和非柠檬酸静脉全血样本一式两份地应用于新系统的两个测试条批次。另外,将血液收集在柠檬酸盐试管中,处理成血浆,并通过手动倾斜试管法使用rTF / 95和ERM-AD149获得PT结果。 CoaguChek XS系统的INR结果与IRP的平均INR的方法比较表明,平均相对偏差为-0.02%至-0.4%,平均绝对相对偏差为6.4-9.6%,准确度大于95% CoaguChek XS INR相对于IRP的平均INR处于+/- 14%到+/- 21.5%的范围内。研究结果证实了两套新CoaguChek XS系统的成功校准,证明了校准概念的有效性,并证明了与IRP相比新系统的准确性。可以根据新系统的INR结果可靠地做出口服抗凝治疗的临床决策。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号