Test-retest variability of intraocular pressure and ocular pulse amplitude for dynamic contour tonometry: a multicentre study.

机译:动态轮廓眼压计的眼压和眼脉冲幅度的重测变异性:一项多中心研究。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

AIMS: To assess the test-retest variability of intraocular pressure (IOP) and ocular pulse amplitude (OPA) measurements utilising dynamic contour tonometry (DCT) and to evaluate possible influential factors. METHODS: The study included 350 consecutive subjects (175 glaucoma, 175 control; one eye per subject) from seven European centres. IOP was measured once with a Goldmann applanation tonometer (GAT) and twice by DCT (DCT1, DCT2) in a randomised sequence. OPA was also recorded for both DCT measurements. Differences (DCT1-DCT2; OPA1-OPA2; GAT-DCT1; GAT-DCT2) were assessed using the t test. The intraclass coefficient of correlation (ICC) and coefficient of variation (CoV) for DCT and OPA were calculated. RESULTS: DCT1 was 0.6+/-1.6 mm Hg higher than DCT2 (p<0.001); OPA1 was 0.1+/-0.7 mm Hg higher than OPA2 (p=0.02). Results were not influenced by randomisation test order. In both glaucoma and normal subjects, DCT and OPA showed ICC>0.90 and >0.76, and CoV=4.8-5.0% and 10.3-10.5%, respectively. DCT1 and 2 were 2.4+/-2.6 and 1.8+/-2.6 mm Hg higher respectively than GAT (p<0.001). DISCUSSION: DCT test-retest variability was almost perfect for IOP and good for OPA. Tonometry measurements with DCT tended to be overestimated compared with GAT.
机译:目的:利用动态轮廓眼压计(DCT)评估眼压(IOP)和眼搏幅度(OPA)测量的重测变异性,并评估可能的影响因素。方法:该研究包括来自七个欧洲中心的350名连续受试者(175名青光眼,175名对照;每名受试者一只眼睛)。用戈德曼压平眼压计(GAT)测量IOP一次,用DCT(DCT1,DCT2)随机测量两次。还记录了两次DCT测量的OPA。使用t检验评估差异(DCT1-DCT2; OPA1-OPA2; GAT-DCT1; GAT-DCT2)。计算了DCT和OPA的类内相关系数(ICC)和变异系数(CoV)。结果:DCT1比DCT2高0.6 +/- 1.6 mm Hg(p <0.001); OPA1比OPA2高0.1 +/- 0.7 mm Hg(p = 0.02)。结果不受随机测试顺序的影响。在青光眼和正常受试者中,DCT和OPA分别显示ICC> 0.90和> 0.76,CoV = 4.8-5.0%和10.3-10.5%。 DCT1和2分别比GAT高2.4 +/- 2.6和1.8 +/- 2.6 mm Hg(p <0.001)。讨论:DCT重测变异性对于IOP几乎是完美的,而对OPA则很好。与GAT相比,使用DCT进行眼压测量的趋势往往被高估。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号