Although a variety of timing techniques and data acquisition strategies have been used for three-dimensional contrast-enhanced MR angiography, many are still limited by inadequate overall reliability, limited spatial resolution, or complexity. A technique is presented in this work in which contrast arrival is detected in the targetted vasculature in real time using MR fluoroscopy. Upon detection the operator triggers a 3D MR angiographic acquisition which uses an elliptical centric view order. It is shown that the view order intrinsically provides a high degree of venous suppression which in turn allows acquisition times of 30 seconds or longer, permitting high spatial resolution. The reliability of fluoroscopic triggering in bolus detection is shown to be over 90%. The technique provides high quality contrast-enhanced MR angiograms for many vascular regions.
机译:尽管已将各种定时技术和数据采集策略用于三维对比度增强的MR血管造影,但许多技术仍受到整体可靠性不足,空间分辨率有限或复杂性的限制。在这项工作中提出了一种技术,其中使用MR荧光镜实时检测目标血管中的对比度到达。一旦检测到,操作员便会触发使用椭圆中心视图顺序的3D MR血管造影采集。结果表明,观察顺序本质上提供了高度的静脉抑制,进而允许了30秒或更长时间的采集时间,从而实现了高空间分辨率。推注检测中荧光镜触发的可靠性显示超过90%。该技术为许多血管区域提供了高质量的对比度增强的MR血管造影照片。