Scan quality effect on glaucoma discrimination by glaucoma imaging devices.

机译:扫描质量对青光眼成像设备对青光眼鉴别的影响。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

AIM: To evaluate, within ocular imaging scans of acceptable quality as determined by manufacturers' guidelines, the effects of image quality on glaucoma discrimination capabilities. METHODS: One hundred and four healthy and 75 glaucomatous eyes from the Advanced Imaging in Glaucoma Study (AIGS) were imaged with GDx-VCC, HRT II and StratusOCT. Quality score (QS>/=8), pixel standard deviation (SD/=5) were used as quality parameter cut-offs, respectively. GDx nerve fibre indicator (NFI) and HRT Moorfields regression analysis (MRA) classifications and OCT mean retinal nerve fibre layer (RNFL) thickness were used as the discriminatory parameters. Logistic regression models were used to model the dichotomous clinical classification (healthy vs glaucoma) as a function of image-quality parameters and discriminatory parameters. RESULTS: Quality parameter covariates were statistically non-significant for GDx and HRT but had an inverse effect on OCT in predicting disease (a higher SS had a lower probability of glaucoma). Age was a significant covariate for GDx and HRT, but not OCT, while ethnicity and interaction between the image quality and the institute where scans were acquired were significant covariates in the OCT models. CONCLUSION: Scan quality within the range recommended as acceptable by the manufacturer of each imaging device does not affect the glaucoma discriminating ability of GDx or HRT but does affect Stratus OCT glaucoma discrimination.
机译:目的:在制造商指南确定的可接受质量的眼科成像扫描中,评估图像质量对青光眼辨别能力的影响。方法:用GDx-VCC,HRT II和StratusOCT对来自青光眼高级影像学研究(AIGS)的104眼健康和75眼青光眼进行了成像。质量得分(QS> / = 8),像素标准偏差(SD / = 5)分别用作质量参数的截止值。 GDx神经纤维指示剂(NFI)和HRT Moorfields回归分析(MRA)分类和OCT平均视网膜神经纤维层(RNFL)厚度用作判别参数。使用逻辑回归模型将二分临床分类(健康与青光眼)建模为图像质量参数和判别参数的函数。结果:质量参数协变量对GDx和HRT在统计学上无统计学意义,但对OCT预测疾病的影响却相反(SS越高,青光眼的可能性越低)。年龄是GDx和HRT的重要协变量,但不是OCT,而种族和图像质量与获取扫描的机构之间的相互作用是OCT模型中的重要协变量。结论:扫描质量在每个成像设备制造商建议的可接受范围内不会影响GDx或HRT的青光眼辨别能力,但会影响Stratus OCT青光眼的辨别力。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号