Improving locoregional hyperthermia delivery using the 3-D controlled AMC-8 phased array hyperthermia system: a preclinical study.

机译:使用3-D控制的AMC-8相控阵热疗系统改善局部热疗的传递:一项临床前研究。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

BACKGROUND: The aim of this study is preclinical evaluation of our newly developed regional hyperthermia system providing 3-D SAR control: the AMC-8 phased array consisting of two rings, each with four 70 MHz waveguides. It was designed to achieve higher tumour temperatures and improve the clinical effectiveness of locoregional hyperthermia. METHODS: The performance of the AMC-8 system was evaluated with simulations and measurements aiming at heating a centrally located target region in rectangular (30 x 30 x 110 cm) and elliptical (36 x 24 x 80 cm) homogeneous tissue equivalent phantoms. Three properties were evaluated and compared to its predecessor, the 2-D AMC-4 single ring four waveguide array: (1) spatial control and (2) size of the SAR focus, (3) the ratio between maximum SAR outside the target region and SAR in the focus. Distance and phase difference between the two rings were varied. RESULTS: (1) Phase steering provides 3-D SAR control for the AMC-8 system. (2) The SAR focus is more elongated compared to the AMC-4 system, yielding a lower SAR level in the focus when using the same total power. This is counter-balanced by (3) a superficial SAR deposition which is half of that in the AMC-4 system, yielding a more favourable ratio between normal tissue and target SAR and allowing higher total power and up to 30% more SAR in the focus for 3 cm ring distance. CONCLUSION: The AMC-8 system is capable of 3-D SAR control and its SAR distribution is more favourable than for the 2-D AMC-4 system. This result promises improvement in clinical tumour temperatures.
机译:背景:这项研究的目的是对我们新开发的提供3-D SAR控制的区域热疗系统进行临床前评估:AMC-8相控阵由两个环组成,每个环带有四个70 MHz波导。它旨在达到更高的肿瘤温度并改善局部热疗的临床效果。方法:通过模拟和测量旨在评估矩形(30 x 30 x 110 cm)和椭圆形(36 x 24 x 80 cm)均匀组织等效体模中位于中心的目标区域,评估了AMC-8系统的性能。对三个特性进行了评估,并将其与其前身2-D AMC-4单环四波导阵列进行了比较:(1)空间控制和(2)SAR焦点的大小,(3)目标区域外最大SAR之间的比率和SAR成为重点。两个环之间的距离和相位差是变化的。结果:(1)相位控制为AMC-8系统提供了3-D SAR控制。 (2)与AMC-4系统相比,SAR焦点要更长一些,在使用相同的总功率时,焦点上的SAR级别会更低。 (3)表面SAR沉积是AMC-4系统的一半,从而抵消了这种情况,从而在正常组织和目标SAR之间产生了更有利的比率,并允许更高的总功率和高达30%的SAR更多聚焦3厘米环距离。结论:AMC-8系统具有3-D SAR控制能力,并且其SAR分布比2-D AMC-4系统更为有利。该结果有望改善临床肿瘤温度。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号