AIM: Some surgeons consider hand held surgical keratometers unreliable. This may be due to incorrect use through not realising that the distance that the keratometer is held from the cornea influences the shape of the image. When a keratometer is held closer to the astigmatic cornea, the elliptical image will appear more circular, particularly for larger degrees of astigmatism. However, the keratoscopic astigmatic ruler (KAR) has design features that correct the hitherto unrecognised problems with the use of a hand held keratometer. This study assesses the reliability and accuracy of measurement of astigmatism using the KAR. METHODS: The KAR and the Bausch & Lomb keratometer (B&L) were compared using six back surface toric cut contact lens blanks representing 1 to 6 dioptres of astigmatism. Two observers (one experienced in the use of the keratometers, the other a novice) took eight randomly repeated "masked" measurements of each lens blank with the KAR and four measurements with the B&L in a similar fashion. RESULTS: There was no difference between the measurements with either instrument by each of the observers (p = 0.95, ANOVA). The standard error of measurement for the KAR was 0.59 D, for the B&L, 0.31 D. The intraclass correlation coefficient of reliability for the KAR was 0.90 and for the B&L it was 0.97. The coefficient of repeatability for the KAR was plus or minus 0.83 D, and for the B&L plus or minus 0.77 D. The interobserver reliability for the KAR was 0.898, and for the B&L, 0.975. CONCLUSION: These results suggest that the KAR has good reliability and reproducibility and compares favourably with the B&L keratometer. Inexperience with use does not affect reliability.
机译:目的:一些外科医生认为手持手术角膜曲率计不可靠。这可能是由于未意识到角膜曲率仪与角膜的距离会影响图像的形状而导致的错误使用。当角膜曲率计保持更靠近散光角膜时,椭圆形图像将显得更圆,尤其是对于较大的散光度。但是,角膜散光像散尺(KAR)的设计特征是使用手持式角膜曲率计纠正了迄今为止无法识别的问题。这项研究评估了使用KAR测量散光的可靠性和准确性。方法:使用代表1至6屈光度散光的六个背面复曲面切割隐形眼镜毛坯比较了KAR和博士伦角膜曲率计(B&L)。两名观察员(一名经验丰富的角膜曲率计,另一名是新手)以类似的方式对每个镜片毛坯使用KAR进行了八次随机重复的“掩盖”测量,并对B&L进行了四次测量。结果:每个观察者使用两种仪器进行的测量之间没有差异(p = 0.95,ANOVA)。 KAR的标准测量误差为0.59 D,B&L的标准误差为0.31D。KAR的组内可靠性相关系数为0.90,B&L的可靠性为0.97。 KAR的可重复性系数为正负0.83 D,B&L的正负系数为负0.77D。KAR的观察者间可靠性为0.898,B&L的观察者间可靠性为0.975。结论:这些结果表明KAR具有良好的可靠性和可重复性,并且与B&L角膜曲率计相比具有优势。使用经验不足不会影响可靠性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号