A comparison of four methods of tonometry: method agreement and interobserver variability.

机译:四种眼压测量方法的比较:方法一致性和观察者间的可变性。

获取原文
获取原文并翻译 | 示例
       

摘要

AIM: To compare the inter-method agreement in intraocular pressure (IOP) measurements made with four different tonometric methods. METHODS: IOP was measured with the Goldmann applanation tonometer (GAT), Tono-Pen XL, ocular blood flow tonograph (OBF), and Canon TX-10 non-contact tonometer (NCT) in a randomised order in one eye of each of 105 patients with ocular hypertension or glaucoma. Three measurements were made with each method, and by each of two independent GAT observers. GAT interobserver and tonometer inter-method agreement was assessed by the Bland-Altman method. The outcome measures were 95% limits of agreement for IOP measurements between GAT observers and between tonometric methods, and 95% confidence intervals for intra-session repeated measurements. RESULTS: The mean differences (bias) in IOP measurements were 0.4 mm Hg between GAT observers, and 0.6 mm Hg, 0.1 mm Hg, and 0.7 mm Hg between GAT and Tono-Pen, OBF, and NCT, respectively. The 95% limits of agreement were smallest (bias +/-2.6 mm Hg) between GAT observers, and larger for agreement between the GAT and the Tono-Pen, OBF, and NCT (bias +/-6.7, +/-5.5, and +/-4.8 mm Hg, respectively). The OBF and NCT significantly underestimated GAT measurements at lower IOP and overestimated these at higher IOP. The repeatability coefficients for intra-session repeated measurement for each method were +/-2.2 mm Hg and +/-2.5 mm Hg for the GAT, +/-4.3 mm Hg for the Tono-Pen, +/-3.7 mm Hg for the OBF, and +/-3.2 mm Hg for the NCT. CONCLUSIONS: There was good interobserver agreement with the GAT and moderate agreement between the NCT and GAT. The differences between the GAT and OBF and between the GAT and Tono-Pen probably preclude the OBF and Tono-Pen from routine clinical use as objective methods to measure IOP in normal adult eyes.
机译:目的:比较用四种不同眼压测量法进行眼内压(IOP)测量的方法间一致性。方法:使用高曼压平眼压计(GAT),Tono-Pen XL,眼血流眼压计(OBF)和佳能TX-10非接触式眼压计(NCT)以每只眼中的每只眼的随机顺序测量IOP。高眼压或青光眼患者。每种方法以及两名独立的GAT观察员均进行了三项测量。通过Bland-Altman方法评估了GAT观察者和眼压计的方法之间的一致性。结果测量为GAT观察者之间和眼压测量方法之间IOP测量的一致性限制为95%,会话内重复测量的95%置信区间。结果:GOP观察者之间的IOP测量值的平均差异(偏差)为0.4 mm Hg,GAT与Tono-Pen,OBF和NCT之间的IOP测量值分别为0.6 mm Hg,0.1 mm Hg和0.7 mm Hg。在GAT观察者之间,一致度的95%最小(偏差+/- 2.6 mm Hg),对于GAT与Tono-Pen,OBF和NCT之间的一致度,偏差最大(偏差+/- 6.7,+ /-5.5,和+/- 4.8 mm Hg)。在较低的IOP下,OBF和NCT明显低估了GAT测量值,而在较高的IOP下,则高估了这些值。每种方法的会话间重复测量的重复性系数对于GAT为+/- 2.2 mm Hg和+/- 2.5 mm Hg,对于Tono-Pen为+/- 4.3 mm Hg,对于GAT笔为+/- 3.7 mm Hg。 OBF,NCT为+/- 3.2 mm Hg。结论:与GAT的观察员之间达成了良好的共识,而NCT与GAT之间的共识也很适中。 GAT和OBF之间的差异以及GAT和Tono-Pen之间的差异可能排除了将OBF和Tono-Pen作为常规方法用于测量正常成年人眼中IOP的客观方法。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号