Reproducibility of ophthalmodynamometric measurements of central retinal artery and vein collapse pressure.

机译:眼球测功法测量视网膜中央动脉和静脉塌陷压力的可重复性。

获取原文
获取原文并翻译 | 示例
           

摘要

BACKGROUND: To assess the reproducibility of ophthalmodynamometric measurements using a new, Goldmann contact lens associated, device allowing biomicroscopic visualisation of the optic disc. METHODS: The prospective clinical study included 87 eyes of 58 subjects presenting with a normal fundus (n=40), or ocular diseases (n=47). With topical anaesthesia, a Goldmann contact lens, fitted with a pressure sensor mounted into the holding ring of the contact lens, was placed onto the cornea. Pressure was applied onto the globe through the contact lens, and the pressure values obtained when the central retinal vessels started pulsating were noted. The measurements were performed 10 times. RESULTS: The mean coefficients of variation for redeterminations of the collapse pressure of the central retinal vein and artery were 16.3% (SD 11.4%), and 8.5% (4.1%), respectively. CONCLUSIONS: A simple and new, Goldmann contact lens associated, ophthalmodynamometer allows central retinal artery and vein collapse pressure measurements which are reproducible in a clinical setting.
机译:背景技术:为了评估使用新的,与戈德曼(Goldmann)接触镜相关联的设备进行眼功测力测量的可重复性,该设备可以对光盘进行生物显微镜可视化。方法:前瞻性临床研究包括58名眼底正常(n = 40)或眼部疾病(n = 47)的受试者的87只眼。在局部麻醉下,将装有压力传感器的戈德曼角膜接触镜安装在角膜接触镜的固定环上。通过隐形眼镜将压力施加到球体上,并记录视网膜中央血管开始搏动时获得的压力值。测量进行了10次。结果:重新确定视网膜中央静脉和动脉的塌陷压力的平均变异系数分别为16.3%(SD 11.4%)和8.5%(4.1%)。结论:一个简单而新颖的,与戈德曼角膜接触镜相关的检眼镜测力计可以测量视网膜中央动脉和静脉的塌陷压力,这在临床环境中是可重现的。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号