首页> 外文期刊>International heart journal >Direct microscopic observation of striations in a fractured section of a sirolimus-eluting stent (Cypher Bx Velocity(R)) indicates induction of stent fracture by continuous shear stress.
【24h】

Direct microscopic observation of striations in a fractured section of a sirolimus-eluting stent (Cypher Bx Velocity(R)) indicates induction of stent fracture by continuous shear stress.

机译:直接显微镜观察西罗莫司洗脱支架(Cypher Bx Velocity(R))的断裂部分中的条纹表明,连续剪切应力导致支架断裂。

获取原文
获取原文并翻译 | 示例
           

摘要

A 73-year-old woman with severe congestive heart failure was treated by implantation with a sirolimus-eluting stent (SES; Cypher Bx Velocity((R))) in the left main coronary artery (LMCA) using the staged T-stent, kissing balloon, and hugging balloon techniques. Follow-up coronary multislice computed tomography after 10 months revealed that SES was completely fractured in 2 directions; the fractured stent appeared in the shape of the letter "L" and had migrated into the aorta. An SES fragment was surgically removed and subsequent electron microscopy revealed striations (striped patterns in fractured sections) on the fracture plane, indicating continuous shear stress after SES implantation in the LMCA. This case provides direct evidence of continuous shear stress on the SES and indicates the necessity of improving the structure of the stent such that it can withstand shear stress.
机译:一位患有严重充血性心力衰竭的73岁妇女通过分阶段T型支架在左主冠状动脉(LMCA)内植入西罗莫司洗脱支架(SES; Cypher Bx Velocity(R))进行治疗,接吻气球和拥抱气球技巧。 10个月后进行的冠状动脉多层计算机断层扫描显示,SES在两个方向上完全断裂。破裂的支架以字母“ L”的形状出现并迁移到主动脉中。通过外科手术切除了SES碎片,随后的电子显微镜检查显示了骨折平面上的条纹(断裂部分上的条纹图案),表明将SES植入LMCA后具有持续的剪切应力。这种情况提供了SES上连续剪切应力的直接证据,并表明需要改进支架的结构以使其能够承受剪切应力。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号