首页> 外文期刊>International heart journal >Temperature-controlled cooled-tip radiofrequency linear ablation of the atria guided by a realtime position management system.
【24h】

Temperature-controlled cooled-tip radiofrequency linear ablation of the atria guided by a realtime position management system.

机译:在实时位置管理系统的指导下,对心房进行温度控制的冷端射频线性消融。

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

Due to the difficulty in producing a transmural linear lesion and the possibility of complications such as thrombus formation leading to thromboembolism, the catheter-based maze procedure remains problematic. We tested, in pigs, the possibility of using a temperature-controlled cooled-tip radiofrequency (RF) ablation system together with a realtime position management (RPM) system to create a transmural linear lesion uncomplicated by thrombus formation.Nine pigs underwent insertion of two electrode catheters (each with two ultrasound electrodes), one into the coronary sinus (CS) and one into the right ventricular apex (references for ultrasound-based non-fluoroscopic three-dimensional mapping). A cooled-tip catheter (with two ultrasound electrodes) was introduced into the right atrium. Linear right atrial ablation was performed with a custom radiofrequency (RF) generator. The catheter was perfused with 0.66 mL/second of saline. RF was delivered for 60 seconds at a target temperature of 40 degrees C. A linear ablation line was created between the superior vena cava and inferior vena cava. Three-dimensional isochronal maps were created during CS pacing before and after ablation. In 4 of the 9 pigs, a transmural linear ablation line was confirmed by three-dimensional mapping and postmortem macroscopic examination. No endocardial thrombus formation was noted. Temperature-controlled cooled-tip RF linear ablation guided by an RPM system appears to have potential for creating linear lesions in the atria. Further studies are needed to determine whether such an ablation technique and the parameters used will facilitate successful completion of the catheter-based maze procedure.
机译:由于难以产生透壁线性病变,并且可能发生诸如血栓形成等导致血栓栓塞的并发症,因此基于导管的迷宫手术仍然存在问题。我们在猪中测试了使用温度控制的冷端射频(RF)消融系统和实时位置管理(RPM)系统来创建透壁性线性病变的可能性,该病变不会形成血栓形成.9头猪接受了两个电极导管(每个带有两个超声电极),一个进入冠状窦(CS),一个进入右心室尖(基于超声的非荧光三维三维映射的参考)。将冷却尖端的导管(带有两个超声电极)引入右心房。使用定制射频(RF)发生器进行线性右房消融。用0.66mL /秒的盐水灌注导管。 RF在40摄氏度的目标温度下输送60秒。在上腔静脉和下腔静脉之间形成了一条线性消融线。在消融前后在CS起搏期间创建了三维等时图。在9头猪中的4头中,通过三维映射和验尸后的宏观检查证实了透壁线性消融线。没有发现心内膜血栓形成。由RPM系统引导的温度控制的冷端RF线性消融似乎具有在心房中形成线性病变的潜力。需要进一步研究以确定这种消融技术和所使用的参数是否将有助于成功完成基于导管的迷宫手术。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号