Although diagnostically indispensable, magnetic resonance imaging (MRI) has been, until recently, contraindicated in patients with an implantable cardiac device. MR conditional cardiac devices are now widely used, but the mode programming needed for safe MRI has yet to be established. We reviewed the details of 41 MRI examinations of patients with a MR conditional device. There were no associated adverse events. However, in 3 cases, paced beats competed with the patient's own beats during the MRI examination. We describe 2 of the 3 specific cases because they illustrate these potentially risky situations: a case in which the intrinsic heart rate increased and another in which atrial fibrillation occurred. Safe MRI in patients with an MR conditional device necessitates detailed MRI mode programming. The MRI pacing mode should be carefully and individually selected.
机译:尽管诊断上必不可少,但直到最近,磁共振成像(MRI)在具有可植入心脏装置的患者中都是禁忌的。 MR条件心脏设备现已得到广泛使用,但是安全MRI所需的模式编程尚未建立。我们回顾了使用MR条件设备进行的41例MRI检查的细节。没有相关的不良事件。但是,在3例中,在MRI检查期间,节律性搏动与患者自身的搏动竞争。我们描述了这3种具体情况中的2种,因为它们说明了这些潜在的风险情况:一种情况是内在心率增加,另一种情况是发生心房颤动。对于具有MR条件设备的患者,要进行安全的MRI检查,必须进行详细的MRI模式编程。 MRI起搏模式应谨慎选择。