Agreement of Immunoassay and TandemMass Spectrometry in the Analysis of Cortisol and Free T4: Interpretation and Implications for Clinicians

机译:免疫分析和串联质谱分析在皮质醇和游离T4分析中的协议:解释及对临床医生的启示

获取原文
获取原文并翻译 | 示例
           

摘要

Objective. To quantify differences in results obtained by immunoassays (IAs) and tandem mass spectrometry (MSMS) for cortisol and free thyroxine (FT4). Design & Patients. Cortisol was measured over 60 minutes following a standard ACTH stimulation test (n = 80); FT4 was measured over time in two cohorts of pregnant (n = 57), and nonpregnant (n = 28) women. Measurements. Samples were analyzed with both IA and MSMS. Results. Results for cortisol by the two methods tended to agree, but agreement weakened over the 60-minute test and was worse for higher (more extreme) concentrations. The results for FT4 depended on the method. IA measurements tended to agree with MSMS measurements when values fell within “normal levels”, but agreement was not constant across trimester in pregnant women and was poorest for the extreme (low/high) concentrations. Correlations between MSMS measurements and the difference between MSMS and IA results were strong and positive (0.411 < r < 0.823; all P < .05). Conclusions. IA and MSMS provide different measures of cortisol and FT4 at extreme levels, where clinical decision making requires the greatest precision. Agreement between the methods is inconsistent over time, is nonlinear, and varies with the analyte and concentrations. IA-based measurements may lead to erroneous clinical decisions.
机译:目的。为了量化通过免疫测定(IAs)和串联质谱(MSMS)获得的皮质醇和游离甲状腺素(FT4)结果的差异。设计与患者。在标准的ACTH刺激试验(n = 80)后的60分钟内测量皮质醇。随着时间的推移,在两组孕妇(n = 57)和未妊娠(n = 28)的女性中测量了FT4。测量。用IA和MSMS分析样品。结果。两种方法得出的皮质醇结果趋于一致,但在60分钟的测试中一致性减弱,并且对于更高(更极端)的浓度而言更差。 FT4的结果取决于方法。当数值落在“正常水平”之内时,IA的测量值往往与MSMS的测量值一致,但是孕妇的妊娠期一致性不是恒定的,并且在极端(低/高)浓度时最差。 MSMS测量值与MSMS与IA结果之间的差异之间的相关性强且呈正相关(0.411

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号