首页> 外文期刊>investigative radiology >Digital Radiography with Photostimulable Storage PhosphorsControl of Detector Latitude in Chest Imaging
【24h】

Digital Radiography with Photostimulable Storage PhosphorsControl of Detector Latitude in Chest Imaging

机译:Digital Radiography with Photostimulable Storage PhosphorsControl of Detector Latitude in Chest Imaging

获取原文
   

获取外文期刊封面封底 >>

       

摘要

RATIONALE AND OBJECTIVES.A widely used digital radiography system based on a photostimulable storage phosphor (PSP) detector was analyzed with regard to radiographic contrast changes that result from the adjustment of detector latitude (x-ray sensitivity range) in the normal processing of chest images.METHODS.Images of an acrylic step wedge were acquired using the digital system in a mode that permitted direct control of effective detector latitude. The images were post-processed in conditions duplicating those used for portable chest examinations, and contrast was measured.RESULTS.Increases in effective detector latitude provided only marginal radiographic contrast gains in the subdiaphragm- equivalent areas of the laser-printed digital film image, while causing large reductions in radiographic contrast in the lung-equivalent region.CONCLUSION.Detector latitude is an important variable that should be monitored or controlled in investigations that compare reader performance using conventional and digital systems.