Theoretical research indicates that, with all variables controlled, Class II biosafety cabinets (BSCs) with brushless direct current (BLDC) motors can save up to 80% in energy costs over conventional cabinets with alternating current (AC) motors. To validate this theory, the University of Michigan (Ann Arbor) conducted a field study on three brands of biosafety cabinets. The study showed that the BLDC cabinets can save significant amounts of energy over conventionally powered units and that this benefit is realized over a wide range of filter loading conditions. Therefore, the energy savings available from BLDC Class II biosafety cabinets warrant earnest consideration before purchasing AC motor Class II biosafety cabinets.
机译:理论研究表明,在所有变量均受控制的情况下,与无刷直流(BLDC)电动机相比,具有无刷直流(BLDC)电动机的II级生物安全柜(BSC)可以节省多达80%的能源成本。为了验证该理论,密歇根大学(安娜堡分校)对三个品牌的生物安全柜进行了实地研究。研究表明,BLDC机柜比传统的动力单元可节省大量能源,并且在各种过滤器负载条件下都可以实现这一优势。因此,在购买交流电动机II级生物安全柜之前,必须认真考虑BLDC II级生物安全柜的节能效果。