Evaluation of the albumin cobalt binding (ACB) assay for measurement of ischaemia-modified albumin (IMA) on the Beckman Coulter LX-20.

机译:在贝克曼库尔特LX-20上评估白蛋白钴结合(ACB)分析以测定缺血修饰的白蛋白(IMA)。

获取原文
获取原文并翻译 | 示例
       

摘要

BACKGROUND: In the presence of ischaemia, albumin undergoes changes resulting in the formation of ischaemia-modified albumin (IMA). Increased serum concentrations of IMA have been found in patients with myocardial ischaemia. The purpose of this study was threefold: to evaluate the albumin cobalt binding (ACB) assay for measurement of IMA on the Beckman Coulter LX-20; to establish a reference range for IMA; and to investigate the relationship between IMA and total albumin concentrations. METHODS: The ACB assay was evaluated under the following headings: imprecision, accuracy and reliability. A reference range was established on a population of 81 healthy subjects. RESULTS: The within-batch coefficient of variation (CV) at IMA concentrations of 88, 99 and 120 KU/L were 1.4, 2.0 and 2.5%, respectively. The between-batch CVs at 74, 84 and 123 KU/L were 3.4, 3.3 and 3.0%, respectively. Comparison with the Cobas Mira Plus showed a mean negative bias of 7 KU/L. The 97.5th percentile established on our reference population was 110 KU/L. A significant inverse relationship was found between total serum albumin and IMA concentrations (r = -0.66, P < 0.0001). Correcting the IMA concentrations for total albumin in our reference population, using a formula devised in this study, yielded a range similar to that of uncorrected IMA. CONCLUSIONS: The ACB assay was found to have acceptable precision and performed very satisfactorily on the Beckman Coulter LX-20. A correction to measured IMA concentrations, to take into account total albumin concentrations, may need to be applied for the proper interpretation of IMA results.
机译:背景:在缺血的情况下,白蛋白发生变化,导致形成缺血修饰的白蛋白(IMA)。在心肌缺血患者中发现IMA的血清浓度升高。这项研究的目的是三方面的:评估用于在贝克曼库尔特LX-20上测量IMA的白蛋白钴结合(ACB)分析;为IMA建立参考范围;并研究IMA与总白蛋白浓度之间的关系。方法:ACB分析按照以下标题进行评估:不精确,准确性和可靠性。在81名健康受试者的人群中建立了参考范围。结果:IMA浓度分别为88、99和120 KU / L时,批内变异系数(CV)分别为1.4%,2.0%和2.5%。 74、84和123 KU / L的批间CV分别为3.4%,3.3%和3.0%。与Cobas Mira Plus的比较显示平均负偏差为7 KU / L。在我们的参考人群中确定的97.5个百分点为110 KU / L。发现血清总白蛋白与IMA浓度之间存在显着的反比关系(r = -0.66,P <0.0001)。使用本研究中设计的公式校正参考人群中总白蛋白的IMA浓度,得出的范围与未校正的IMA相似。结论:发现ACB测定具有可接受的精度,并且在Beckman Coulter LX-20上进行的结果非常令人满意。考虑到总白蛋白浓度,可能需要对测得的IMA浓度进行校正,以正确解释IMA结果。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号