A feed-through microwave power measuring system has been established at NPL at 10 0 GHz using a thin-film standard bolometer mount of effective efficiency 96.50#xb1;0.37. The value of Kc#x2019;, the calibration factor of the feed through system, obtained is 10.695. The limits of total uncertainty in the measurement ofKe, are +0.70 and#x2013;0.87. The technique employed in the measurement ofKeand the error-flow chart are also given. This feed-through power measuring system is very useful for determining the effective efficiency/calibration factor of unknown bolometer mounts, and for the calibration of microwave power meters at 10.0 GHz upto 10 W.