首页> 外文期刊>Annals of clinical biochemistry. >Validation of a liquid chromatography tandem mass spectrometry assay for serum creatinine and comparison with enzymatic and Jaffe methods.
【24h】

Validation of a liquid chromatography tandem mass spectrometry assay for serum creatinine and comparison with enzymatic and Jaffe methods.

机译:液相色谱串联质谱法测定血清肌酐的有效性,并与酶法和Jaffe方法进行比较。

获取原文
获取原文并翻译 | 示例
           

摘要

BACKGROUND: Glomerular filtration rate (GFR) can be estimated using creatinine clearance or calculated formulae, methods which rely on creatinine quantification in serum. Creatinine is most frequently measured using the Jaffe method, but this is prone to interference by, for example, bilirubin, protein and ketones. Recent amendments to this assay have been made by manufacturers in an attempt to compensate for the protein interference. METHODS: We developed a novel liquid chromatography tandem mass spectrometry (LC-MS/MS) method for the analysis of creatinine in serum, using a strong cation exchange column to give a short run time of 1.5 min and minimal ion suppression. The method had a lower limit of quantification of 5 micromol/L and the intra- and inter-assay imprecision was <5% and <7%, respectively. Serum samples (n = 182) were subsequently analysed by enzymatic, compensated Jaffe and LC-MS/MS methods and results compared. RESULTS: The compensated Jaffe and enzymatic methods gave similar results tothe LC-MS/MS method at concentrations below 150 micromol/L. At concentrations above this value, creatinine was underestimated by both the compensated Jaffe and enzymatic methods compared with the LC-MS/MS method. CONCLUSIONS: We have developed a novel, simple, sensitive LC-MS/MS assay for use in comparative studies or when the results of either Jaffe or enzymatic assays are in question.
机译:背景:肾小球滤过率(GFR)可以使用肌酐清除率或计算公式来估算,这些方法依赖于血清中肌酐的定量测定。肌酐最常使用Jaffe方法进行测定,但是容易受到例如胆红素,蛋白质和酮的干扰。制造商已对该测定法进行了最新修正,以试图补偿蛋白质干扰。方法:我们开发了一种新型的液相色谱串联质谱法(LC-MS / MS),用于分析血清中的肌酐,方法是使用强阳离子交换柱,以在1.5分钟内实现短运行时间并最小化离子抑制。该方法的定量下限为5 micromol / L,测定内和测定间不精密度分别为<5%和<7%。随后通过酶促,补偿的Jaffe和LC-MS / MS方法分析血清样品(n = 182),并比较结果。结果:当浓度低于150 micromol / L时,补偿的Jaffe和酶法得到的结果与LC-MS / MS方法相似。在浓度高于该值的情况下,与LC-MS / MS方法相比,补偿的Jaffe方法和酶促方法均低估了肌酐。结论:我们已经开发出一种新颖,简单,灵敏的LC-MS / MS测定法,用于比较研究或当Jaffe或酶法测定的结果有疑问时。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号