Measurement of Distensibility of Blood Vessels Using Cineangiograms

机译:使用电影血管造影法测量血管扩张性

获取原文
获取外文期刊封面目录资料

摘要

We developed a new, cineangiographic method to accurately measure the dynamic changes in the internal diameter of human arteries in vivo. Cine films were digitized at a spatial resolution of 4 μm/pixel, using a line image sensor. The vessel edges, with a Gaussian fit to a unilateral profile curve of the vessel, were determined with the aid of a computer program. We measured contrast-filled cylinder vessel models (2 to 7 mm in diameter) and evaluated precision, accuracy and linearity of the diameter measurement. A pulsatile vessel model of about 3.9 mm in internal diameter was used to examine the reliability of our method for detecting arterial wall motion. If the coefficient of variation of the vessel diameter determined cineangiographically was less than 2.2 we considered the cineangiograms sufficiently accurate to determine the internal vessel diameter and evaluate arterial distensibility.
机译:我们开发了一种新的电影血管造影方法,可以准确测量体内人体动脉内径的动态变化。使用线图像传感器以 4 μm/像素的空间分辨率对电影胶片进行数字化。借助计算机程序确定容器边缘,其高斯拟合与容器的单边轮廓曲线相吻合。我们测量了对比填充的圆柱体容器模型(直径为 2 至 7 毫米),并评估了直径测量的精度、准确度和线性度。使用内径约为 3.9 mm 的脉动血管模型来检验我们检测动脉壁运动方法的可靠性。如果通过电影血管造影法确定的血管直径变异系数小于2.2%,我们认为电影血管造影足够准确,可以确定血管内径并评估动脉扩张性。