首页> 外文期刊>neurourology and urodynamics >New method of measuring uroflow in the rat bladder
【24h】

New method of measuring uroflow in the rat bladder

机译:测量大鼠膀胱尿流的新方法

获取原文

摘要

AbstractUrodynamic evaluation of bladder function requires the simultaneous measurement of detrusor pressure and uroflow. Although the rat is a common animal model to study several bladder dysfunctions, including partial outlet obstruction, one of the major problems is that the bladder capacity has been too small for the measurement of uroflow. We have developed a new method of measuring uroflow in the rat that is sensitive at low bladder volumes. The method is based on the change in conductance of the gap between two silver electrodes in a measuring chamber as the chamber fills with saline during voiding. An AC voltage supply drives a current through the saline. As the saline level rises in the chamber the conductance increases, with a corresponding increase in current flow. A linear correlation between time rate of change of electrical current and fluid flow rate exists over a wide range of values and suggests that this new rat flowmeter is both accurate and reproducible. © 1992 Wiley‐Liss, I
机译:摘要膀胱功能的尿动力学评估需要同时测量逼尿肌压力和尿流。虽然大鼠是研究几种膀胱功能障碍(包括部分出口梗阻)的常见动物模型,但主要问题之一是膀胱容量太小,无法测量尿流。我们开发了一种测量大鼠尿流的新方法,该方法在低膀胱容积时敏感。该方法基于测量室中两个银电极间隙的电导变化,因为该室在排空过程中充满盐水。交流电源驱动电流通过盐水。随着腔室中盐水含量的升高,电导增加,电流也相应增加。电流的时间变化率与流体流量之间存在线性相关性,在很宽的值范围内,这表明这种新型大鼠流量计既准确又可重复。© 1992 威利·利斯,我

著录项