Use of Gray Values in CBCT and MSCT Images for Determination of Density: Influence of Variation of FOV Size

机译:使用CBCT和MSCT图像中的灰度值确定密度:FOV大小变化的影响

获取原文
获取原文并翻译 | 示例
           

摘要

Purpose: To determine the effect of field of view (FOV) size on gray values in cone beam computed tomography (CBCT) and multislice computed tomography (MSCT) images. Material and Methods: A phantom made up of 3 cylinders containing distilled water, plaster, and motor oil was constructed and inserted into an acrylic cylinder filled with distilled water. The phantom was scanned with a CBCT and MSCT device using 3 FOV sizes. Gray value of each material was evaluated in 40 axial slices, and the comparison between the results obtained with the same FOV size was made.
机译:目的:确定视场(FOV)大小对锥束计算机断层扫描(CBCT)和多层计算机断层扫描(MSCT)图像中灰度值的影响。材料和方法:制作一个由3个装有蒸馏水,石膏和机油的圆柱体组成的模型,并将其插入装有蒸馏水的丙烯酸圆柱体中。用CBCT和MSCT设备使用3种FOV尺寸扫描体模。在40个轴向切片中评估每种材料的灰度值,并对在相同FOV尺寸下获得的结果进行比较。