摘要

Introduction: Atrial synchronous left ventricular (LV) only pacing using two leads and VDD pacemaker could be a cost effective alternative to conventional cardiac resynchronization therapy (CRT).Methods: We implanted right atrial (RA) and LV leads with VDD pulse generator (LV only pacing) in five carefully screened heart failure patients who could not afford conventional CRT. All had NYHA class III/IV symptoms despite maximal guideline directed medical therapy. The sensed atrioventricular delay was programmed to pre-excite the LV and achieve fusion beat Response to treatment was assessed at 6 months.
机译:简介:仅使用两根导线和VDD起搏器进行心房同步左心室(LV)的起搏可能是传统心脏再同步治疗(CRT)的一种经济有效的方法。方法:我们将VDD脉冲发生器(LV仅起搏)在五名经过仔细筛查的,无法负担常规CRT的心力衰竭患者中。尽管有最大指导方针的药物治疗,所有患者均具有NYHA III / IV级症状。对感觉到的房室延迟进行编程,以预先激发LV并达到融合搏动。在6个月时评估对治疗的反应。