首页> 外文期刊>artificial organs >Hemodynamic Evaluation of a New Left Ventricular Assist Device: An Integrated Cardioassist Catheter as a Pulsatile Left Ventricle‐Femoral Artery Bypass
【24h】

Hemodynamic Evaluation of a New Left Ventricular Assist Device: An Integrated Cardioassist Catheter as a Pulsatile Left Ventricle‐Femoral Artery Bypass

机译:新型左心室辅助装置的血流动力学评估:作为搏动性左心室股动脉搭桥术的集成心脏辅助导管

获取原文

摘要

Abstract:We developed a new left ventricular (LV) assist catheter (transaortic valve from LV to femoral artery) that is implemented percutaneously and exerted as a synchronous pulsatile partial LV flow support with conventional intraaortic balloon pump (IABP) driving system and centrifugal pump system. We investigated whether this pulsatile LV‐aorta bypass is superior to the IABP from a hemodynamic viewpoint. Ten dogs with profound heart failure were placed on this device. and hemodynamic measurements were performed under on‐off study of this system. The results revealed a significant increase of mean aortic pressure, total cardiac output, and myocardial blood flow, and a significant reduction of left atrial (LA) pressure and LV afterload estimated by diastolic pressure‐time indexitension‐time index (DPTI/TTI) measurement compared with baseline values and also with IABP exertion alone. These findings suggest that this system is of clinical value for supporting an impaired LV that is intractable under IABP counterpu
机译:摘要:我们开发了一种新的左心室(LV)辅助导管(从左心室到股动脉的经主动脉瓣),该导管通过经皮实施,并作为同步搏动性部分左心室血流支持,与传统的主动脉内球囊泵(IABP)驱动系统和离心泵系统一起发挥作用。从血流动力学的角度来看,我们研究了这种搏动性左心室主动脉搭桥术是否优于 IABP。十只患有严重心力衰竭的狗被放置在该设备上。并在该系统的开关研究下进行血流动力学测量。结果显示,与基线值相比,平均主动脉压、总心输出量和心肌血流量显著增加,舒张压-时间指数-时间指数 (DPTI/TTI) 测量估计的左心房 (LA) 压力和左心室后负荷显着降低,也与单独使用 IABP 运动相比。这些发现表明,该系统对于支持在 IABP 对位下难以治疗的受损 LV 具有临床价值

著录项

获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号