Failures of the RM finger prosthesis joint replacement system

机译:Failures of the RM finger prosthesis joint replacement system

获取原文
获取原文并翻译 | 示例
           

摘要

In our unit a high failure rate of the RM finger prosthesis joint replacement system was noted, prompting a review of cases. A series of patients underwent implantation under the care of one surgeon and the results were monitored. Twenty-one devices were implanted of which 16 were inserted for rheumatoid disease. Patients were reviewed regularly and the implant performance was assessed critically along with survival of the implant to revision, infection or death of the patient. The mean follow-up was 32 months. Unacceptable failure rates at early and medium term stages were identified, with 15 of the implants revised by 2 years. Loosening was the commonest mode of failure. The authors do not recommend the use of this implant, especially in cases of rheumatoid arthritis.