The method of standard additions with electrochemically generated spikes was elaborated and demonstrated on examples of O-2 and H-2 chromatographic analysis in standard mixtures. While oxygen portions may be generated directly in a stream of inert gas by solid-state electrolysis, a standard hydrogen-containing mixture is introduced into the electrochemical cell followed by controlled electrochemical consumption of hydrogen. In both cases the analysis characteristics are similar: the recovery is 90-120 and exhibits a slight sensitivity towards the flow rate of the analyzed oxygen-containing gas. The relative error associated with the non-ideal proportionality between the spike and chromatograph signal is within 10-15 for operation in the optimum concentration region. The major factors affecting the accuracy of the analysis are associated with detector characteristics, particularly influenced by the sample dilution, precision of flow controllers and linearity of the function "signal vs. spike." The lowest errors of determination were found to be for the analyte content range 1-5 while at oxygen and hydrogen concentrations below 0.5 the interference of the signal associated with other components may lead to overestimated results.

著录项

获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号