首页> 外文期刊>Biomedical Engineering: Applications, Basis and Communications >MULTIMODALITY COMPARATIVE STUDY: QUANTITATIVE ASSESSMENT OF MRI CONTRAST AGENT DOSE REDUCTION
【24h】

MULTIMODALITY COMPARATIVE STUDY: QUANTITATIVE ASSESSMENT OF MRI CONTRAST AGENT DOSE REDUCTION

机译:多模态比较研究:MRI对比剂剂量减少的定量评估

获取原文
获取原文并翻译 | 示例
           

摘要

Contrast agents are used in 30-40% of diagnostic MRI scans. These agents are expensive and cause adverse effects in some patients. One way to reduce the amount of contrast agent needed would be to improve the signal-to-noise ratio (SNR). We hypothesized that pre-magnetizing samples of the paramagnetic contrast agent gadodiamide before scanning would increase the SNR. Samples of gadodiamide in saline were magnetized with a permanent magnet before being placed in a phantom and scanned in either a closed 1.5 T MRI or an open 0.35 T MRI. Samples were scanned 0-5 hours after magnetization. Samples of gadodiamide that had not been magnetized were also scanned. In the 1.5 T instrument, we found an increase in SNR in pre-magnetized samples that was highest three hours after magnetization, up to a 1.8-fold increase at low concentrations of gadodiamide. The improvement in SNR was more modest in the low-field MRI scanner, peaking two hours after magnetization. This relatively simple method has the potential to reduce the amount of contrast agents needed for diagnostic MRI scans.
机译:造影剂用于诊断性MRI扫描的30-40%。这些药物价格昂贵,并且对某些患者造成不良影响。减少所需造影剂数量的一种方法是提高信噪比(SNR)。我们假设在扫描之前将顺磁性造影剂gadodiamide样品预磁化会增加SNR。将盐水中的gadodiamide样品用永磁体磁化,然后放置在体模中,并在封闭的1.5 T MRI或开放的0.35 T MRI中进行扫描。磁化后0-5小时扫描样品。还对未磁化的gadodiamide样品进行了扫描。在1.5 T仪器中,我们发现预磁化样品中的SNR升高,在磁化后三小时达到最高,在低浓度的gadodiamide时达到1.8倍。在低场MRI扫描仪中,SNR的改善更为适度,在磁化后两个小时达到峰值。这种相对简单的方法具有减少诊断MRI扫描所需的造影剂数量的潜力。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号