首页> 外文期刊>Atmospheric chemistry and physics >Technical Note: Measuring condensation sink and ion sink of atmospheric aerosols with the electrical low pressure impactor (ELPI)
【24h】

Technical Note: Measuring condensation sink and ion sink of atmospheric aerosols with the electrical low pressure impactor (ELPI)

机译:技术说明:使用低压电子冲击器(ELPI)测量大气气溶胶的冷凝槽和离子槽

获取原文
获取原文并翻译 | 示例
           

摘要

We investigate the suitability of ELPI for condensation sink and ion sink measurements. The aim is to find the simple calibration factors by which the measured ELPI current can be converted to condensation or ion sinks. The calibration is based on DMPS and ELPI measurements within the period 15??"25 May 2005 at a boreal forest site in Southern Finland. The values of condensation sink and ion sink were calculated from the DMPS size distributions using their theoretical definitions. After that the values were compared to theoretical and measured ELPI current, and calibration factors were specified. For condensation sink the calibration factor was found to be 7.27E-06 s~(-1)fA~(-1) and for ion sink 8.55E-06 s~(-1)fA ~(-1). Simply by multiplying the total current of the outdoor ELPI by these factors, the values of condensation sink and ion sink can be measured.
机译:我们调查了ELPI对于冷凝池和离子池测量的适用性。目的是找到简单的校准因子,通过这些校准因子,可以将测得的ELPI电流转换为冷凝或离子阱。该校准基于2005年5月15日至25日在芬兰南部的一个北方森林站点上的DMPS和ELPI测量。冷凝水沉和离子水沉的值是使用DMPS尺寸分布的理论定义计算得出的。将该值与理论和测量的ELPI电流进行比较,并指定校准系数,对于冷凝槽,校准系数为7.27E-06 s〜(-1)fA〜(-1),对于离子槽为8.55E- 06 s〜(-1)fA〜(-1)。只需将室外ELPI的总电流乘以这些因素,即可测量出冷凝阱和离子阱的值。