首页> 外文期刊>Analytical and bioanalytical chemistry >Defining a system suitability limit to decide on column deterioration and to facilitate column transfers in chiral supercritical fluid chromatography
【24h】

Defining a system suitability limit to decide on column deterioration and to facilitate column transfers in chiral supercritical fluid chromatography

机译:Defining a system suitability limit to decide on column deterioration and to facilitate column transfers in chiral supercritical fluid chromatography

获取原文
获取原文并翻译 | 示例
           

摘要

The separation of enantiomers is an important requirement during the entire drug life cycle in the pharmaceutical industry. High-performance liquid chromatography and supercritical fluid chromatography (SFC) are the main chromatographic techniques used to separate enantiomers. Since chiral stationary phases are often extensively used once a method has been developed, columns will age and must be replaced after a certain period. However, no practical guidelines exist to determine when a column is deteriorated or to decide whether a transfer to another column (with the same chiral selector) is successful. In this study, a system suitability limit for resolution was defined, based on an intermediate (time-different) precision study in SFC on four immobilized polysaccharide-based columns that only differed in manufacturer or particle size. This system suitability limit could be used to decide on column deterioration or as a requirement to evaluate whether a separation transfer was successful. Some method adaptations may be necessary to obtain successful transfers. An approach was proposed, which helped the analyst to make successful transfers.

著录项

获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号