Correlation of iCare ic100 tonometry with iCare TA01i in screening of unselected population in Northern Finland Birth Cohort Eye study

机译:Correlation of iCare ic100 tonometry with iCare TA01i in screening of unselected population in Northern Finland Birth Cohort Eye study

获取原文
获取原文并翻译 | 示例
       

摘要

Purpose: iCare tonometers are easy-to-use and portable devices for measuring the intraocular pressure (IOP). Purpose was to evaluate the IOP values measured by both novel iCare ic100 and conventional model TA01i devices in unselected population. Methods: IOP was measured with iCare ic100 and TA01i tonometers in 149 participants aged 32–33?years (born in 1985 or 1986) of the Northern Finland Birth Cohort Eye 2 study. The right eye of each participant was selected for analysis. We also collected data on axial length, corneal curvature and central corneal thickness (CCT). Bland-Altman plot was used for comparing the values obtained by these devices. Results: Mean IOP measured with the ic100 device was 13.8 (3.4) mmHg, with TA01i it was 12.5 (3.0)?mmHg. The mean difference between these devices was 1.30?mmHg ( p ?