An in vitro evaluation of the accuracy of Dentaport ZX apex locator in enlarged root canals.

机译:Dentaport ZX根尖定位器在扩大的根管中的准确性的体外评估。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

BACKGROUND: The aim of this study was to examine the effects of root canal irrigants on the accuracy of Dentaport ZX electronic apex locator (EAL) in enlarged root canals. Our previous study revealed that as the diameter of the root canal increased, the electronically measured length with small size files became shorter in the presence of blood. It is not known whether different canal irrigants would interfere with the reading accuracy of an EAL in enlarged root canals. METHODS: A total of 45 extracted single-rooted human teeth were used. In Stage 1, canals were instrumented using #10-40 K-files with a #40 K-file as the master apical file (MAF). The teeth were randomly divided into five groups and mounted in an experimental apparatus. The following irrigants were used during electronic canal measurements: Group A = 0.5% NaOCl; Group B = 2.5% NaOCl; Group C = 15% EDTA; Group D = 0.8% chlorhexidine (CHX); and Group E = RC Prep. In Stage 2, the canals were enlarged using a #60 K-file as the MAF. In Stages1 and 2, the apical portion of the canals was instrumented using the step-back sequence (up to a #80 K-file). In Stage 3, the canals were enlarged again using a #80 K-file as the MAF. In each stage, the canal length was measured with a Dentaport ZX using #10 and #40, #10 and #60, and #10 and #80 K-files for Stages 1, 2 and 3, respectively. Data were analysed by two-way ANOVA and Fisher's PLSD test. RESULTS: Statistical analysis showed significant differences among all groups (p < 0.01). There was no significant difference between Group A and B at each stage and between Group D and E at Stage 2. A statistical significant difference was observed between #10 K-file and other files (#40, #60 and #80) at each stage (p < 0.01). In Groups A and B, there was no significant difference between #10 and #40 K-files (Stage 1) and between #10 and #60 K-files (Stage 2). CONCLUSION: The Dentaport ZX was accurate and not adversely affected by the presence of 0.5% or 2.5% NaOCl and EDTA in the enlarged canals, and the measuredlengths obtained with small and large size files were comparable. However, it was accurate in the presence of CHX and RC Prep only when large size files were used, and the length measured with small size files was greater than the actual length.
机译:背景:这项研究的目的是检查根管冲洗液对扩大的根管中Dentaport ZX电子根尖定位器(EAL)准确性的影响。我们以前的研究表明,随着根管直径的增加,在有血的情况下,电子测量的长度较小的锉变得越来越短。尚不清楚不同的根管冲洗液是否会干扰扩大的根管中EAL的读取精度。方法:共提取45根单根人牙。在第1阶段,使用#10-40 K锉刀和#40 K锉刀作为根尖锉(MAF)来测量运河。将牙齿随机分为五组,并安装在实验设备中。在电子管测量中使用以下冲洗液:A组= 0.5%NaOCl; B组= 2.5%NaOCl; C组= 15%EDTA; D组= 0.8%洗必太(CHX);和E组= RC准备。在第2阶段,使用#60 K锉作为MAF扩大了运河。在第1阶段和第2阶段,使用后退顺序(直至#80 K锉)对根管部分进行器械检查。在第3阶段,使用#80 K锉刀作为MAF,再次扩大了运河。在每个阶段,分别使用#10和#40,#10和#60以及#10和#80 K档的Dentaport ZX测量运河的长度,分别用于阶段1、2和3。通过双向方差分析和Fisher的PLSD测试分析数据。结果:统计分析显示所有组之间存在显着差异(p <0.01)。 A和B组在每个阶段之间以及D和E组在第二阶段之间都没有显着差异。在#10 K文件和其他文件(#40,#60和#80)之间观察到统计学显着差异。阶段(p <0.01)。在A组和B组中,#10和#40 K文件(阶段1)之间和#10和#60 K文件(阶段2)之间没有显着差异。结论:Dentaport ZX准确无误,且在扩大的运河中存在0.5%或2.5%NaOCl和EDTA并没有不利影响,并且无论大小文件,测量的长度都是可比的。但是,只有在使用大尺寸文件的情况下,在CHX和RC Prep的存在下它才是准确的,并且使用小尺寸文件测量的长度大于实际长度。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号