Radiation Methods for Monitoring Parameters of X-Ray Diagnostic Apparatuses

机译:X射线诊断仪参数监测的辐射方法

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

According to SanPiN 2.6.1.1192-03 (Appendix 10) [10], new models of X-ray apparatuses should be regularly tested for conformity of the following parameters of X-ray apparatuses with normative documentations:1. total filtration of the working X-ray beam;2. error of anode voltage settings and first half-value layer;3. pulsation and wave shape of anode voltage;4. error of anode current settings and charge;5. errors of exposure time settings;6. reproducibility of X-ray dose values for photographic examination in manual and automatic modes;7. linear dependence of X-ray radiation dose on charge or exposure time;8. radiation yield, mGy-m2/rnA-rnin, at nominal anode voltage and focal distance 1 m.
机译:根据SanPiN 2.6.1.1192-03(附录10)[10],应定期对新型号的X射线设备进行以下测试:X射线设备的以下参数是否符合规范性文件规定:1.。完全过滤工作的X射线束; 2。阳极电压设置和前半值层误差; 3。阳极电压的脉动和波形; 4。阳极电流设置和充电误差; 5。曝光时间设置错误; 6。 X射线剂量值在手动和自动模式下照相检查的再现性; 7。 X射线辐射剂量对电荷或曝光时间的线性关系; 8。在标称阳极电压和焦距为1 m时的辐射产量mGy-m2 / rnA-rnin。