To evaluate the feasibility and accuracy of continuous monitoring of drag with a fiber optic intravascular sensor in animal model. Methods: An accurate optical design was used to enhance the intensity of light from a 100-microon optic fiber and the fluorescence signal can be detectedThe vary quenching means the vary of the concentration of daunorubicin in rabbit's blood. Daunombicin is determined by sensor based on the feature fluorescence. In a simple animal model, the carotid artery was catheterized with a cannula housing 100-micron optic fiber. Results: The average recovery of all the tested compounds within the set concentration range was 99.9 ~ 108.0%, the within-day repeatability value were acceptable between 2.5~5.3% respectively. The method permitted detection limits as low as 0.05mug·ml~(-1) at a signal -to -noise ratio of 3. Conclusions: Fiber-optic sensor is potentially useful for monitoring blood drug and metabolites in biomedical field.
机译:为了评估在动物模型中用光纤血管内传感器连续监测阻力的可行性和准确性。方法:采用精确的光学设计来增强100微米光纤的光强度,并可以检测荧光信号。不同的猝灭意味着兔血液中柔红霉素浓度的变化。道诺霉素是由传感器根据特征荧光确定的。在简单的动物模型中,将颈动脉插入装有100微米光纤的套管。结果:在设定浓度范围内,所有被测化合物的平均回收率为99.9〜108.0%,日内重复性值分别在2.5〜5.3%之间是可以接受的。该方法在信噪比为3的情况下允许检测限低至0.05mug·ml〜(-1)。结论:光纤传感器可能在生物医学领域用于监测血液药物和代谢产物。