首页> 外文期刊>therapeutic drug monitoring >Comparison of Three Immunoassays for the Determination of Nhyphen;Acetylprocainamideperiod; Determination of False Elevations by a Polyclonal Antibodyhyphen;Based Method
【24h】

Comparison of Three Immunoassays for the Determination of Nhyphen;Acetylprocainamideperiod; Determination of False Elevations by a Polyclonal Antibodyhyphen;Based Method

机译:Comparison of Three Immunoassays for the Determination of Nhyphen;Acetylprocainamideperiod; Determination of False Elevations by a Polyclonal Antibodyhyphen;Based Method

获取原文
           

摘要

The performance of the ROCHE FP NAPA reagent, a newly reformulated monoclonal antibody-based fluorescence polarization immunoassay (FPIA) for TV-acetylprocainamide, was compared with two commercially available polyclonal antibody-based immunoassays, the Abbott TDx and Syva EMIT. Although excellent correlation was observed between the ROCHE FP NAPA reagent and the Syva EMIT reagent, a positive bias was observed in the correlation between the ROCHE FP NAPA reagent and the Abbott TDx reagent. The bias in the TDx method was investigated further by the concurrent high-performance liquid chromatography (HPLC) analysis of patient samples. Insignificant differences (p 0.05) were observed between the HPLC and ROCHE FP values. The TDx system, however, yielded significantly higher (p 0.0005) results than did HPLC.