摘要

A method has been established for the determination of selenium in human serum samples, applying transversely heated graphite furnace AAS with Zeeman-effect background correction. For determination, the serum samples were diluted (1+2) with O.1% Triton X-100 and 0.1% nitric acid. A Pd/Mg (NO_3)_2 matrix modifier was used to stabilize the analyte. To avoid any buildup of carbonate(ms residues on the graphite platform, air ashing wax applied at elevated temperatures. Tile method was evaluated by the analysis of a certified reference material and a comparison study for several samples which was provided by a certified medical laboratory. The detection limit was calculated to 2.0 #mu#g/L Se, i.e., 20 pg.
机译:已经建立了一种采用赛曼效应背景校正的横向加热石墨炉原子吸收光谱法测定人血清中硒的方法。为了确定,将血清样品用0.1%Triton X-100和0.1%硝酸稀释(1 + 2)。 Pd / Mg(NO_3)_2基体改性剂用于稳定分析物。为避免在石墨平台上残留碳酸盐(ms残留物),在高温下使用空气灰化蜡。通过对认证参考材料的分析和由认证医学实验室提供的几种样品的对比研究,评估了平铺方法。计算出的检出限为2.0#mu#g / L Se,即20 pg。